The output power (at full and half settings) versus load resistance (per IEC/EN 60601-2-2,
subclause 6.8.3) is given in the graphs below:
Output Power vs. Load Resistance, Tan Patient Cable Receptacle
Output Power vs. Load Resistance, Black Patient Cable Receptacle
Note: at 50 Ohms, full power setting of the black patient cable receptacle, the current limiting circuit of the
instrument suppressed the output, and thus the output power registered a low value.
The following plot illustrates the relationship between instrument set points and output RMS
voltage.