Error Messages
Fault/Error
Fault Description
LCD Message
Error
Both CUT/COAG buttons pressed
simultaneously
E1: Cut and Coag
simultaneously activated
Error
Wand disconnected while RF active
E2: Wand disconnected
Error
Use-limiting time expired
E3: Wand error
Error
Shorted wand
E4: Wand short
Error
Re-used wand detected
E5: Wand error
Error
Open T/C circuit
E6: Open T/C
Error
Re-used wand detected
E7: Wand error
Error
Controller powered on with re-used wand
E8: Wand error
Fault
Integrity of stored program compromised
(calculated CRC differs from loaded value)
F1: Hardware failure
Fault
Integrity of RAM space compromised (RAM
test failed)
No LCD message can be
displayed
Fault
COP watchdog timer not strobed
F3: Hardware failure
Fault
Output of the DC-DC converter out of
specification
F4: Hardware failure
Fault
Analog measurement failure (measured
value out of specification)
F5: Hardware failure
Fault
Footswitch, fingerswitch or front panel
switch pressed during power on
F6: Switch stuck on
Fault
Over temperature of internal electronics
F7: Internal overtemp