Setup
Synergy
RF™
User’s Guide
DFU-0221-6r0_fmt_en-US
Page 32 of 58
W A R N I N G !
Use only Arthrex approved accessories. Other accessories may result in
increased emissions or decreased immunity of the system. Contact your Arthrex
representative for a complete list of accessories.
DO NOT
modify any
accessory. Failure to comply may result in patient and/or operating room staff
injury.
W A R N I N G !
Do
not
use in the presence of flammable anesthetics or oxidizing gases such as
nitrous oxide, oxygen, or endogenous gases. All oxygen connections must be
leak free for the duration of the surgical proce
d
ure
.
4.6
Setting up a Footswitch
Only use footswitches developed by Arthrex specifically for the
AR-9800 Synergy
RF
console.
NOTES: Setup for a footswitch is the same for all models. The console detects
which version is attached and allows appropriate functions.
The footswitch connector cannot be properly inserted into the console
receptacle if the corresponding black dots are not aligned.
The pins of the footswitch connector may be damaged if the black dots
are not aligned before the footswitch is fully inserted.
The user may experience intermittent operation of the footswitch if the
connector is not fully seated into the console receptacle.
Insert the footswitch connector into the console’s footswitch receptacle. Align the
black dots on the connector and receptacle so they engage easily.
4.7
Setting up the Arthrex Single Use, disposable Apollo
RF®
device (Probe)
W A R N I N G !
Check that the disposable RF device packaging and sterile barrier is intact
before opening the product and introducing it to the sterile surgery field.
If the packing has been compromised DO NOT use the product.
During the electrosurgical procedure, the patient should not be allowed to
come into direct contact with grounded metal objects such as surgical
table frame, instrument table, etc. which may lead to patient injury.
When not in use, place the Probe in a clean, dry, non-conductive and highly
visible area not in contact with the patient. Inadvertent activation while in
contact with the patient may cause burns.
Содержание Synergy AR-9800
Страница 2: ...This page intentionally left blank ...
Страница 25: ...Technical Specifications SynergyRF User s Guide DFU 0221 6r0_fmt_en US Page 23 of 58 3 6 Power Curves ...
Страница 26: ...Technical Specifications SynergyRF User s Guide DFU 0221 6r0_fmt_en US Page 24 of 58 ...
Страница 27: ...Technical Specifications SynergyRF User s Guide DFU 0221 6r0_fmt_en US Page 25 of 58 ...
Страница 28: ...Technical Specifications SynergyRF User s Guide DFU 0221 6r0_fmt_en US Page 26 of 58 ...
Страница 29: ...Technical Specifications SynergyRF User s Guide DFU 0221 6r0_fmt_en US Page 27 of 58 ...
Страница 30: ...Technical Specifications SynergyRF User s Guide DFU 0221 6r0_fmt_en US Page 28 of 58 ...