DrillSaw Highspeed 200™
User’s Guide
DFU-0225-2 Rev. 0
Page 41 of 58
Always place the main power switch in the “Off (O), position” and disconnect
the power before cleaning the AR-200C console.
Always comply with the instructions issued by the manufacturer of
the cleaning disinfectant regarding concentration, exposure times,
temperature, and material compatibility
.
NEVER allow the console receptacles to have any contact with
liquids. If there is dust or moisture on the receptacles, remove with
dry compressed air. ONLY dry connectors should be plugged into
the console.
Do NOT clean the device with abrasive cleaning or disinfectant
compounds, solvents, or other materials that could scratch or
damage the device.
Do not use machine (automatic) cleaning on the AR-200C console
6.3
Motor with Cable (AR-200M, AR-200M-ISO)
and Spray Clip (AR-200SP)
The motor with cable (AR-200M, AR-200M-ISO) is supplied
non-sterile
.
After cleaning, sterilize the motor.
Clean and disinfect the motor with cable immediately after every treatment!
The Spray Clip is provided
non-sterile
.
The Spray Clip can be cleaned and sterilized with the suggested cleaning devices
provided by Arthrex:
•
AR-200 - CLEANING ADAPTER FOR SPRAY CLIP
•
AR-200 – SILICONE TUBING FOR CLEANING
ADAPTER, 10 PIECES
Clean the Spray Clip before each use.
Please use dedicated cleaning adapters and tubing defined as
accessories in this User’s Guide, for cleaning and sterilization of
AR-200SP Spray Clip.
Prolonged soaking in cleaning agents may deteriorate the
components of the device. Always comply with the instructions
issued by the manufacturer of the cleaning disinfectant regarding
concentration, exposure times, temperature, and material
compatibility.