DrillSaw Mini 300™ System
User’s Guide
DFU-0223-2 Rev. 0
Page 2
12.
Use only
Arthrex approved accessories. Other accessories may result in
increased emissions or decreased immunity of the system. Contact your Arthrex
representative for a complete list of accessories.
DO NOT
modify any accessory.
Failure to comply may result in injury to the patient and/or operating room staff.
13. Do not use in the presence of flammable anesthetics or oxidizing gases such as
nitrous oxide, oxygen, or endogenous gases. All oxygen connections must be
leak free for the duration of the surgical procedure.
14. After autoclaving, the accessory devices are VERY HOT. Handle with care to
avoid burns.
15. The equipment is
NOT
suitable for use in potentially explosive atmospheres or
with potentially explosive mixtures of anesthetic substances containing oxygen
or nitrous oxide.
16. Z one M is defined as a “medical environment” and constitutes the part of a room
in which potentially explosive atmospheres may form due to the use of
anesthetics or medical antiseptics and antibacterial soaps; such atmospheres
are typically localized and temporary. Zone M comprises a truncated pyramid
below the operating table which is tilted outwards at a 30° angle.
17. Zone G, also known as an “enclosed medical gas system”, does not necessarily
include areas enclosed around all sides, in which explosive mixtures are
continuously or temporarily generated, directed or used in small quantities.
18. Only fit an attachment/accessory when the handpiece is at a complete stop and
in safe position. Safe position is annotated by the “S” on the mode switch.
19. Do not interfere with the rotary instruments or saws while they are in motion.
20. Never touch an attachment or press the release mechanism of the collet while
the device is in motion.
21. To avoid potential injury, never allow a rotary or saw attachment to make contact
with the surrounding soft tissues and vessels.
22. Never use the Arthrex DrillSaw Mini 300 System in areas near strong magnetic
fields.
23. Do not place the handpiece on the patient.
24. Disconnect the rechargeable battery immediately after using the handpiece.
25. Only use accessories which are in proper working order. Damaged or improperly
functioning accessories must be sent in for service.
26. It is recommended to have a backup system available should the need arise.
The PRECAUTION! symbol identifies methods and procedures that
must be followed to avoid damaging the device or causing it to
malfunction.
Содержание DrillSaw Mini 300 System
Страница 2: ...This page intentionally left blank ...
Страница 43: ...DrillSaw Mini 300 System User s Guide DFU 0223 2 Rev 0 Page 37 This page intentionally left blank ...
Страница 46: ...DrillSaw Mini 300 System User s Guide DFU 0223 2 Rev 0 Page 40 This page intentionally left blank ...
Страница 47: ...DrillSaw Mini 300 System User s Guide DFU 0223 2 Rev 0 Page 41 This page intentionally left blank ...