950-0073-00 B SM, English, Arthrex Synergy UHD4 System
41 of 74
12. The Surgical monitor will display one of the following.
When the White Balance has been completed successfully, a Green Check Mark with
WHITE BALANCE below will be shown on screen.
Figure 19 - White Balance OK
When the White Balance has not been completed successfully, a Red X with WHITE
BALANCE below will be shown on screen.
Figure 20 - White Balance Fail
13. If the White Balance Operation has been successful, the camera will enter the Surgical Ready
Mode and be ready for surgical operation.
14. If the White Balance Operation has not been successful, you must move the Tip of the
Endoscope closer or farther from the White Gauze until the operation can be completed
successfully.
15. Once the White Balance Operation has been successfully completed, the Camera Head buttons
will function as defined on the Synergy
UHD4
Initial Screen Figure .