4400 Series
LaserSource
User’s Manual · Page 59
Error Messages
Error
Code
Description
Cause
E-100
General Error
The error code is non-specific, and is generally
used when no other error code is suitable.
E-102
Message too long
The message is too long to process (USB/Serial
only).
E-123
Path not found
The message used an invalid path command
(USB/Serial only).
E-124
Data mismatch
The message contained data that did not match
the expected format (USB/Serial only).
E-126
Too few or too many
elements
The command requires more or less than the
number of parameters actually supplied.
E-127
Change not allowed
An attempt was made to change a parameter
that cannot be changed, or is currently read-
only.
E-128
Delay or active script
terminated
An executing script, or a *WAI or DELAY
function was terminated before completing.
E-201
Data out of range
The message attempted to set a value that was
outside the allowable range (USB/Serial only).
E-202
Invalid data type
When trying to parse the message, the data was
in an invalid format (USB/Serial only).
E-204
Suffix not valid
An invalid number base suffix (radix) was
encountered when parsing a number
(USB/Serial only).
E-217
Configuration Recall
failed
An attempt to recall a configuration failed. This
can be caused if no configuration exists in the
selected slot, the slot number is out of range, or
if the configuration is corrupted.
E-218
Configuration Save
failed
An attempt to save a configuration failed. This
can be caused if the slot number is out of range,
or the configuration memory is corrupted.
E-220
Script Save Failed
An attempt to save a script failed. This can be
caused if the script number is out of range, or
the script memory is corrupted.
E-221
Cannot embed script
A script was executed that contained a reference
to another script.
E-222
Cannot execute script
An attempt to execute a script failed. This can be
caused if the script number is out of range, no
script exists for the selected index, or the script
memory is corrupted.
E-501
Interlock shutdown
output
The interlock input (pins 1 and 2 of the output
connector) were not shorted when the output
was on.
E-504
Laser current limit
disabled output.
The laser output was turned off because a
current limit was detected and the
corresponding bit in the OUTOFF register was
set.