4400 Series
LaserSource
User’s Manual · Page 21
Menu
Description
Factory
Default
Cable R
The cable resistance, in ohms. This setting can
be used to compensate for voltage losses in the
cable due to cable and connector resistances.
Cable R
is ignored in Vf (AVC) Mode. For more
information about this feature, see
Using the
Cable R Setting
section. Not available when
using remote voltage sense.
0.0000
Ω
Int Contact
Enables or disables intermittent contact. Not
available in
Io (Pulse)
,
Io (Ext Trig)
, or
Io
(Burst)
modes
Off
PD Resp
This factor is used by the unit to convert from
monitor photodiode current into optical power.
The value is in terms of microamps per milliwatt
(
μ
A/mW), such that power = photodiode current
divided by the factor.
0.00
μ
A/mW
PD Bias
This is the photodiode bias voltage, which is
applied to the PD+/PD- pins of the output
connector.
5.0V
On Delay
The delay, from the time the
Output
button is
pressed to when the output is actually
energized.
3000ms
Aux Sensor Menu
ThermA
The A term in the Steinhart-Hart equation for
use with auxiliary sensors.
1.12924E-03
ThermB
The B term in the Steinhart-Hart equation for
use with auxiliary sensors.
2.34108E-04
ThermC
The C term in the Steinhart-Hart equation for
use with auxiliary sensors.
0.87755E-07
T# Lo Lim
Low temperature limit for Aux Sensor 1, 2 or 3
-99°C
T# Hi Lim
High temperature limit for Aux Sensor 1, 2 or 3
125°C
R# Lo Lim
Low resistance limit for Aux Sensor 1, 2 or 3
0.01k
Ω
R# Hi Lim
High resistance limit for Aux Sensor 1, 2 or 3
45k
Ω
Digital I/O Menu
Di1 Func
Selects the action to perform when Digital Input
1 is true.
None
Di1 Invert
Inverts the active state of Digital Input 1.
No
Di2 Func
Selects the action to perform when Digital Input
2 is true.
None
Содержание LaserSource 4400 Series
Страница 1: ......
Страница 64: ...Copyright 2020 Arroyo Instruments All Rights Reserved P N 530 1055C...