www.armano-messtechnik.com • p. 9 of 20 • GB • 09/22 • B75 •
B08-590
Operating Instructions
PAXd – Display
5.3 Calibration
(only if write protection is disabled)
The device can be process-calibrated at two points to match a given input signal. A low input signal (not necessar-
ily 0 %) is applied and the actual value is entered. Then a high input signal (not necessarily 100 %) is applied and
the actual value is entered. If you accept the calibration (by selecting Yes in the save menu), the device will operate
according to the new settings. You can reset the device to factory calibration points by selecting reset calibration
in the calibration menu. Please note that the programming mode will not time out while performing a calibration.
Sensor 2
Calibration
Txt 53
Sensor 1
Sensor 2
*5
CJC sensor
Loop Curr
*26
2
Do Calib
S2 Calib
Txt 54
Reset Calib
Do Calib
2
Do Calib
Loop Calib
Txt 54
Reset Calib
Do Calib
2
3
*5
Set S2 Low
Txt 74
0.0
S2 Cal Low
Txt 55
-200°C
850°C
2
*27
*14
4.00
Calib 4 mA
Txt 55
3.50 mA
23.00 mA
20.00
Calib 20 mA
Txt 56
3.50 mA
23.00 mA
2
2
*12
*13
Do Calib
S1 Calib
Txt 54
Reset Calib
Do Calib
2
Do Calib
CJC Calib
Txt 54
Reset Calib
Do Calib
2
Set S1 Low
Txt 74
0.0
S1 Cal Low
Txt 55
-200°C
850°C
2
*27
*14
Set CJC Low
Txt 74
0.0
CJC Cal Low
Txt 55
-40.0
135.0
2
*27
*14
3
3
3
3
3
3
3
3
Set S2 High
Txt 75
*27
100.0
S2 Cal High
Txt 56
-200°C
850°C
2
*14
Set S1 High
Txt 75
*27
100.0
S1 Cal High
Txt 56
-200°C
850°C
2
*14
Set CJC High
Txt 75
*27
100.0
CJC Cal Hi
Txt 56
-40.0
135.0
2
*14
*26
7501V111-UK
31
Calibration
(only if Write-protection is disabled)
The device can be process-calibrated in two points to fit a given input signal. A low input signal (not necessarily 0%) is applied
and the actual value is entered. Secondly a high input signal (not necessarily 100%) is applied and the actual value is entered.
If you accept to use the calibration (by selecting Yes in the Save menu) the device will work according to the new adjustments.
You can return the device to factory calibration points by selecting Reset Calibration from the Calibration menu. Please note
that while performing a calibration the programming mode will not time out.
Only shown if dual sensor function is
selected.
*5
*13
20 mA fixed output configured internally.
4 mA fixed output configured internally.
*12
Default value is current input value.
No timeout in this menu.
*14
*26
Not possible if Loop Current is not active.
*27
No timeout in this menu.
Not available in
HART 5 mode
Sensor 2
Loop Curr
Sensor 1
CJC sensor
Reset Calib
Do Calib
Reset Calib
Do Calib
Reset Calib
Do Calib
Reset Calib
Do Calib
“Monitor”
“Monitor”
“Monitor”
“Monitor”
“Monitor”
“Monitor”
“Monitor”
“Monitor”
Only shown if dual sensor function is selected.
4 mA fixed output configured internally.
20 mA fixed output configured internally.
Default value is current input value. No timeout in this menu.
Not possible if loop current is not active.
No timeout in this menu.
Not available in
HART 5 mode