MAINTENANCE & CLEANING
General instructions
The exterior of the device can be cleaned as needed. The interior of the device should be
cleaned during the process of exchanging ice blocks. Do not clean the interior when it is in use
for vaccine storage, as doing so will cause unnecessary temperature exchange and thus reduce
vaccine hold time.
To clean the exterior of the device
You can clean the exterior of the device when it is loaded with vaccines, however, do not shake
or tilt the device. You must avoid damaging the vaccine vials.
1.
Use soap and water to wash the device with a hose, sponge, or cloth. Do not use
abrasives or cleaning powders.
Note: The device’s electronics are can tolerate some moisture. However, do not
submerge the device.
2.
Dry the device with a lint-free towel or allow it to air dry.
To clean the interior of the device
:
To clean the interior of the Arktek, take these steps:
1.
Remove the cap.
2.
Remove the vaccine stacks. Follow established cold chain vaccine handling practices
with any removed vaccines.
3.
If needed, clean the vaccine cup-stack containers with soap and water and allow them
to dry.
4.
Remove and set aside the ice blocks.
5.
Use mild soap and water or a 1% chlorine bleach solution to clean the inside of the
device with a hose, sponge, or cloth. Do not scrub the interior with abrasives or cleaning
powders.
Caution: Take care of the wiring that is affixed on the interior of the device; these wires
are for the sensors that monitor the internal temperature of the device.
Caution: The device should not be submerged in water.
6.
Thoroughly wipe down the interior of the device with a towel. The interior of the device
can be air-dried if time allows. You can also turn the device upside down as needed to
allow it to drain.
7.
Leave the cap off of the device during storage for next use. It will remain tethered to the
device.