b
4-3-2
Cleaning the Probe
Chapter 4 Maintenance
4-06 ■ SE-1520
Correctly insert the check plate so that the arrows on the check plate face
forward.
Prepare the check plate.
1.
The analyzer reads the bar
c o d e o f t h e c h e c k p l a t e
automatically.
●
S t a r t C h e c k M E A S
S t o p( S T O P )
After the bar code is read, the
p l a t e m o v e s t o t h e o p t i c a l
block automatically and the
check measurement starts.
Then check results are printed.
●
M e a s u r i n g C H K
0 2 0
S t o p( S T O P )
If an error occurs, clean the point of contact shown in the next page.
e.g.: Normal printing
e.g.: Abnormal printing
Select [3. Check Meas] on the
Maintenance Menu.
Set the check plate and press
[START] key.
●
●
For probe cleaning, perform measurements with the check plate.
If an error occurs after the measurement, clean the probe according to the
following procedure.