Chapter 5 Troubleshooting

When a problem occurs in the instrument, an error code will be displayed as 

shown below.

If an error code is displayed, refer to the table below to take appropriate 

action. If this does not solve the problem and you were using a mobile 

battery, connect the AC adapter and check operation. If you were using the 

AC adapter from the beginning or if the problem still occurs after connecting 

the AC adapter, contact your distributor.

NOTE:

If action taken to address an error or trouble may impact performance, 

perform a control measurement as necessary.

5.1 If a Warning Occurs

5.1.1 From Warning Occurrence to Remedy

The instrument notifies you of a warning by:

• Emitting short beeps.

• Displaying a warning code and message.

If a warning occurs, take the necessary actions to remove the cause (see 

“5.1.2. Causes and Remedies”). If the warning persists, turn off the power 

and contact your distributor.

5.1.2 Causes and Remedies

5.2 If an Error Occurs

5.2.1 From Error Occurrence to Remedy

The instrument notifies you of an error by:

• Emitting short beeps.

• Displaying an error code and message.

If an error occurs, take the necessary actions to remove the cause (see 

“5.2.2. Causes and Remedies”). If the error persists, turn off the power and 

contact your distributor.

Remedy

No.

Cause

5.3 If Trouble Occurs

5.3.1 From Trouble Occurrence to Remedy

The instrument alerts you of a trouble by:

• Emitting short beeps.

• Displaying a trouble code.

If same as above occurs, take the necessary actions to remove the cause 

(see “5.3.2. Causes and Remedies”). If the trouble persists, turn off the power 

and contact your distributor.

5.3.2 Causes and Remedies

A problem occurred in 

communication with an 

external device.

Connect the external device with the 

communication cable correctly.

E-102

The date and time 

settings are incorrect.

Set the date and time correctly.

E-103

The photometric window 

is dirty.

Clean the photometric window.

E-122

The ambient temperature 

of the instrument is less 

than 10°C or over 35°C.

Adjust ambient temperature so as to 

create a suitable environment for 

measurement (10 - 35°C).

E-123

The reagent strip has 

expired.

Use a new reagent strip and repeat 

measurement.

E-124

The tray was withdrawn 

during measurement.

Press the     button to clear the error, 

and repeat measurement.

E-126

The measurement count 

limit of the inserted 

IC-CHIP has been 

exceeded.

Insert the IC-CHIP inside of the box 

of the reagent strip to be measured 

into the port on the back of the 

instrument, and repeat measure-

ment.

E-128

Unable to read the 

reagent strip barcode.

Press the     button to clear the error, 

and repeat measurement.

E-129

The photometric window 

is dirty.

Press the     button to clear the error, 

and use a cleaning pad to clean the 

photometric window. The cleaning 

method is indicated in the leaflet that 

is included with the cleaning pad.

E-130

The measurement unit’s 

light intensity is abnormal.

Adjust the light intensity gain.

E-131

A different IC-CHIP is 

inserted.

Press the     button to clear the error, 

insert the IC-CHIP inside of the box of 

the reagent strip to be measured into 

the dedicated port on the back of the 

instrument, and repeat measurement.

E-132

IC-CHIP is not inserted.

Insert the IC-CHIP inside of the box 

of the reagent strip to be measured 

into the IC-CHIP insertion port.

NO IC

QC measurement has not 

been performed, or the 

expiration date of the QC 

measurement has 

passed.

Perform QC measurement.

PCS

The strip tray is not 

inserted.

Insert the strip tray. (The warning will 

be automatically cleared.)

TRAY 

OUT

The strip tray is still 

inserted.

Remove the strip tray. (The warning 

will be automatically cleared.)

TRAY IN

The strip tray was not 

inserted within 10 

seconds of pressing the 

    button.

Press the     button and repeat 

measurement from the beginning.

A reagent strip is still in 

the strip tray.

Withdraw the strip tray, discard the 

reagent strip, and insert the strip tray 

again.

T OVER

RGNT 

IN

A malfunction occurred 

with the optical unit.

Turn off the power and contact your 

distributor.

T-201

A malfunction occurred 

with the internal 

temperature control.

Adjust ambient temperature so as to 

create a suitable environment for 

measurement (10 - 35°C).

T-221

An internal malfunction 

occurred.

Turn off the power and contact your 

distributor.

T-222

A malfunction occurred 

with the optical unit.

Clean the transmission photometric 

holes or photometric window.

If a used reagent strip is still in the 

strip tray, remove it.

T-223

An internal malfunction 

occurred.

Turn off the power and contact your 

distributor.

T-999

Содержание Lab 004

Страница 1: ...in close proximity to sources of strong electromagnetic radiation as these may interfere with proper operation Premise Symbols Read this operating manual thoroughly before using the instrument This operating manual gives an overview of the instrument and the proper procedures for operation and maintenance Follow the instructions in this operating manual in order not to defeat the purpose of the pr...

Страница 2: ...d in package insert of reagent Measurement principle Listed in package insert of reagent Dual wavelength transmitted light measurement Number of reagent strips measurable at once 1 Measurement time Listed in package insert of reagent Approximately 4 μL Sample consumption Listed in package insert of reagent 1 sample Approximately 10 minutes in environments 25 C and below Memory capacity 50 measurem...

Страница 3: ...adapter Use a battery with a rating of 5 V and 2 0 2 4 A Do not disassemble the instrument Do not modify the instrument Disassembly and modification of the instrument may result in exposure to pathogenic microorganisms or cause fire or damage to the instrument and consequently lead to personal injury 1 4 2 Precautions for Instrument Relocation Before relocating the instrument read the following no...

Страница 4: ... infectious diseases Improper handling of blood may cause infection to the user or other individuals by pathogenic microorganisms Dispose of used samples in accordance with local regulations for biohazardous waste When using a sample that has been stored in a refrigerator or freezer allow the sample to acclimate to the ambient temperature 10 35 C prior to measurement Do not store a sample in direc...

Страница 5: ... from a New Box Wear disposable gloves to prevent exposure to pathogenic microorganisms If sample adheres to the instrument wipe it off with a cotton swab or gauze dampened with distilled water Make sure that no dirt is left and then ensure that the surface is disinfected with alcohol Leaving the instrument with the sample adhered to it may cause infection to the user or other individuals by patho...

Страница 6: ...d time are set a On the standby screen press and hold for 3 seconds to display the MENU screen 1 0 0 will appear in the sub screen with the 1 digit blinking b Press the button The 1 will stop blinking and the middle 0 digit will change to 1 and start blinking c Press the button once The middle digit will change to 2 d Press the button The MENU No will be finalized as 1 2 and the screen will transi...

Страница 7: ...red Use a new reagent strip and repeat measurement E 124 The tray was withdrawn during measurement Press the button to clear the error and repeat measurement E 126 The measurement count limit of the inserted IC CHIP has been exceeded Insert the IC CHIP inside of the box of the reagent strip to be measured into the port on the back of the instrument and repeat measure ment E 128 Unable to read the ...

Страница 8: ...omatic dry biochemical analyzer RATING MADE IN PHILIPPINES Environment During Transport Semi Auto matic dry biochemical analyzer RATING MADE IN PHILIPPINES Storage Environment Storage Environment Environment During Transport This product conforms to Regulation EU 2017 746 In vitro diagnostic medical device Manufacturer Date of manufacture Authorised Representative in the European Community Importe...

Отзывы: