IMPORTANT:

Insert the end of the reagent strip into the opening at the end of the groove 

on the strip tray and then fit the entire strip into the groove. If the reagent 

strip is warped and/or fails to fit into the groove, it may become jammed 

inside the instrument or cause inaccurate measurement results.

2.2.3 Measuring the Sample

a

 Firmly and securely attach the pipette tip to the end of the pipette.

b

 Aspirate the sample with the pipette.

c

 Press the [    ] button, and then dispense a drop of sample on the reagent 

strip. Dispense the sample onto the sample dispense area of the reagent 

strip.

     Press the pipette button all the way in slowly.

NOTE:

Dispense sample carefully to avoid air bubbles.

d

 After you dispense a drop of sample, immediately press the [    ] button and 

insert the strip tray into the instrument. (Insert the strip tray within 10 

seconds of pressing the [    ] button in step 

d

.)

e

 After the measurement time elapses, the measurement results appear on 

the display.

f

 Check the measurement results.

g

 Withdraw the strip tray and take out the used reagent strip.  

Check the reagent strip, and check that the sample has flowed in correctly 

and that there are no large air bubbles or foreign matter in the reaction cell. 

After checking, discard the reagent strip.

h

 The standby screen will appear 5 seconds after withdrawing the strip tray.

2.2.4 Measuring the control

a

 On the standby screen, press the [>] button. The control icon will appear.

b

 Dispense a drop of the control onto the reagent strip, and then perform the 

same procedure as when measuring a sample.

c

 The measurement results are saved as a control measurement.

2.2.5 Understanding results

After measurement, the following information appears on the screen.

Upper line:The item name and measurement number appear alternately. 

If you press the     button, the display changes to the date and 

time.

If there is a status indication (*), the status appears instead of the 

measurement number.

Lower line:The measurement result value and units appear. However, if the 

status is UNDER or OVER, the measurement range appears.

(*) Status

UNDER  Under the lower limit of the measurement range.

Check the conditions. It is possible that drop dispensing failed and 

the sample did not enter the reaction cell, or a large bubble 

formed.

OVER 

Over the upper limit of the measurement range.

2.1.4 Precautions for Tips

IMPORTANT:

 Do not touch tip end with your bare hands. Contamination on the tip end 

may lead to inaccurate measurement results.

 Do not reuse the tip. The tip has a water repellent finish that may come 

off when washed and consequently lead to inaccurate measurement 

results.

2.2 Sample Measurement

Measure samples using a reagent strip. Be sure to start with the 

procedure in “2.2.1 Before Using Reagent Strips from a New Box”.

• Wear disposable gloves to prevent exposure to pathogenic 

microorganisms.

• If sample adheres to the instrument, wipe it off with a cotton 

swab or gauze dampened with distilled water. Make sure that 

no dirt is left, and then ensure that the surface is disinfected 

with alcohol. Leaving the instrument with the sample adhered to 

it may cause infection to the user or other individuals by 

pathogenic microorganisms.

• Dispose of used samples, tips, reagent strips and disposable 

gloves in accordance with local regulations for biohazardous 

waste.

• Be careful not to spill samples.

• Alcohol is very flammable. Keep it away from flames or 

electrical sparks. Handle with caution. Adequately ventilate the 

room during use.

2.2.1 Before Using Reagent Strips from a New Box

Load the reagent information from the IC-CHIP that is included with the 

reagent. You only need to perform this step once when you start using a 

reagent strip from a new box. Before using a reagent strip from a newly 

opened box, be sure to load IC-CHIP on the box.

a

 Insert the IC-CHIP into the IC-CHIP insertion port on the back of the 

instrument.

b

 Press and hold the [>] button for 1 second.

"READING" appears. After the reagent information is read in from the 

IC-CHIP, you will return to the standby screen. If reading of the reagent 

information fails, "NO IC" will appear. Press [>] to exit the display, and then 

reinsert the IC-CHIP and repeat reading.

■ 

When the IC-CHIP of the Urea/BUN reagent is to be read for 

the first time

a

 Insert the IC-CHIP. If you press and hold the [>] button for 1 second, the 

IC-CHIP can be read.

b

 

When the IC-CHIP of the Urea/BUN reagent is to be read for the first time, 

an item selection screen (“Urea” or “BUN”) appears.

c

 Switch to the item you want to use with the [>] button, and then set the item 

name with the [    ] button.

d

 “READING” is displayed. The reading process is completed when the 

display returns to “STANDBY”.

*The item selection screen does not appear when the IC-CHIP is read for the 

second time and thereafter.

The set item name is selected automatically, and the processing shifts to 

d

 

after the operation described in 

a

 is performed.

To change the item name, refer to “When the item name of the Urea/BUN 

reagent is to be changed”.

 

When the item name of the Urea/BUN reagent is to be 

changed

a

 Insert the IC-CHIP. If you press and hold the [>] button for 1 second, 

“READING” is displayed.

b

 

If you press the [<] button while “READING” is being displayed, the item 

selection screen (“Urea” or “BUN”) appears.

c

 Switch to the item you want to use with the [>] button, and then set the 

item with the [    ] button.

d

 “READING” is displayed. The reading process is completed when the 

display returns to “STANDBY”.

2.2.2 Setting the Reagent Strip

a

 Tear open the foil pack at the notch.

b

 Take a reagent strip out without touching the reaction cell.

c

 Set the reagent strip in the groove of the strip tray with the reaction cell 

facing upward.

OK

NG

Dispensing area

Dispensing area

Reaction cell

Base film

Содержание Lab 004

Страница 1: ...in close proximity to sources of strong electromagnetic radiation as these may interfere with proper operation Premise Symbols Read this operating manual thoroughly before using the instrument This operating manual gives an overview of the instrument and the proper procedures for operation and maintenance Follow the instructions in this operating manual in order not to defeat the purpose of the pr...

Страница 2: ...d in package insert of reagent Measurement principle Listed in package insert of reagent Dual wavelength transmitted light measurement Number of reagent strips measurable at once 1 Measurement time Listed in package insert of reagent Approximately 4 μL Sample consumption Listed in package insert of reagent 1 sample Approximately 10 minutes in environments 25 C and below Memory capacity 50 measurem...

Страница 3: ...adapter Use a battery with a rating of 5 V and 2 0 2 4 A Do not disassemble the instrument Do not modify the instrument Disassembly and modification of the instrument may result in exposure to pathogenic microorganisms or cause fire or damage to the instrument and consequently lead to personal injury 1 4 2 Precautions for Instrument Relocation Before relocating the instrument read the following no...

Страница 4: ... infectious diseases Improper handling of blood may cause infection to the user or other individuals by pathogenic microorganisms Dispose of used samples in accordance with local regulations for biohazardous waste When using a sample that has been stored in a refrigerator or freezer allow the sample to acclimate to the ambient temperature 10 35 C prior to measurement Do not store a sample in direc...

Страница 5: ... from a New Box Wear disposable gloves to prevent exposure to pathogenic microorganisms If sample adheres to the instrument wipe it off with a cotton swab or gauze dampened with distilled water Make sure that no dirt is left and then ensure that the surface is disinfected with alcohol Leaving the instrument with the sample adhered to it may cause infection to the user or other individuals by patho...

Страница 6: ...d time are set a On the standby screen press and hold for 3 seconds to display the MENU screen 1 0 0 will appear in the sub screen with the 1 digit blinking b Press the button The 1 will stop blinking and the middle 0 digit will change to 1 and start blinking c Press the button once The middle digit will change to 2 d Press the button The MENU No will be finalized as 1 2 and the screen will transi...

Страница 7: ...red Use a new reagent strip and repeat measurement E 124 The tray was withdrawn during measurement Press the button to clear the error and repeat measurement E 126 The measurement count limit of the inserted IC CHIP has been exceeded Insert the IC CHIP inside of the box of the reagent strip to be measured into the port on the back of the instrument and repeat measure ment E 128 Unable to read the ...

Страница 8: ...omatic dry biochemical analyzer RATING MADE IN PHILIPPINES Environment During Transport Semi Auto matic dry biochemical analyzer RATING MADE IN PHILIPPINES Storage Environment Storage Environment Environment During Transport This product conforms to Regulation EU 2017 746 In vitro diagnostic medical device Manufacturer Date of manufacture Authorised Representative in the European Community Importe...

Отзывы: