Within Run Precision
*Bloodavg.
39 mg/dL
SD = 1.9 mg/dL
*Bloodavg.
82 mg/dL
SD = 2.7 mg/dL
*Bloodavg.
145 mg/dL
CV = 2.9%
*Bloodavg.
179 mg/dL
CV = 3.9%
*Bloodavg.
341 mg/dL
CV = 2.5%
Total Precision
*Controlavg.
32 mg/dL
SD = 1.7 mg/dL
*Controlavg.
121 mg/dL
CV = 3.6%
*Controlavg.
354 mg/dL
CV = 4.3%
Precision:
Precision studies were performed in a laboratory
using the GLUCOCARD Shine System.
Alternative Sites Testing Evaluation
• Comparison of results using various AST sites with the results
of YSI measurements
Alternative site test results for glucose concentration < 75 mg/dL
This study shows that there could be variation of up to 4.3%.
Fingertip
Forearm
Palm
Thigh
Calf
Within ± 5 mg/dL
5/5 (100%)
5/5 (100%)
5/5 (100%)
4/5 (80%)
3/5 (60%)
Within ± 10 mg/dL
5/5 (100%)
5/5 (100%)
5/5 (100%)
5/5 (100%)
5/5 (100%)
Within ± 15 mg/dL
5/5 (100%)
5/5 (100%)
5/5 (100%)
5/5 (100%)
5/5 (100%)