Repeat the same procedure for the
lancing device (step
1
to step
6
).
*
After the pre-cleaning and disinfection procedure, the
control solution should be tested to confirm that the meter
works properly before using the meter. Control solution
tests should be performed with two different levels of
solutions (Control Level 1 and 2). Verify that the test results
are within the range printed on the test strip vial. See pages
19-20 for how to do a control solution test.
7
Note:
If any of the following deterioration signs appear after
pre-cleaning or disinfecting, please stop using the system
and contact Customer Service, 800.566.8558, 24 hours a
day, 7 days a week, US and Canada.
• When the inscriptions on the exterior of the meter (or
lancing device) have been removed.
• When the color of the meter (or lancing device) has
changed or faded.
• When cracks or roughness develop on the meter (or
lancing device).
• When a part of the segment on the meter display does not
flash.
• When control solution test results do not fall within the
stated range on the test strip vial.