Checking the System
The GLUCOCARD Shine XL Meter and GLUCOCARD
Shine Test Strips should be checked using
GLUCOCARD Shine Control Solution, available in two
levels (Level 1 and 2). The GLUCOCARD Shine Control
Solutions contain known amounts of glucose and are
used to check that the meter and test strips are working
properly.
The test strip vials have GLUCOCARD Shine Control
Solution ranges printed on their labels. Compare the
result displayed on the meter to the control solution range
printed on the test strip vial.
Before using a new meter or new vial of test strips, you should
conduct a control solution test using GLUCOCARD Shine
Control Solutions.
Follow the procedure that begins on page 21.
NOTE
• Use GLUCOCARD Shine Control Solutions only.
• Check the expiration date printed on the bottle. When the
control solution bottle is first opened, record the discard
date (date opened plus three months) in the space provided
on the label.
• Make sure the meter, test strips and control solutions are at
room temperature before testing. Control solution tests must
be done at room temperature (68–77 ºF, 20–25 ºC).
• Before using the control solution, shake the bottle, discard
the first one or two drops and wipe the tip clean.
• Close the control solution bottle tightly and store at a
temperature between 46–86 ºF (8–30 ºC).