MAGE1 User Guide
20
V1.5 12-1-2022
9.12. Info
Modbus
register
Modbus
address
No. of
bytes
Data type
No. of
decimal
Min
Value
Max Value
Description
Read/
Write
1000
999
4
Longint
0
-
-
Time
R
1002
1001
4
Longint
0
-
-
Date
R
1004
1003
4
Longint
0
-
-
Unit No.
R
1006
1005
4
Longint
0
-
-
Error (min) R
1008
1007
4
Longint
0
-
-
OK (min)
R
1010
1009
4
Longint
0
-
-
Diameter
R
1012
1011
4
Longint
2
-
-
FirmWare
No.
R
1014
1013
4
Longint
3
-
-
Flow Qn
R
1016
1015
4
Longint
0
-
-
Excitation
counter
R
Time
- time is stored in BCD format HHMMSS (ie 08:33:15 = 0x00083315)
Date
- date is stored in BCD format YYYYMMDD (ie 25.03.2010 = 0x20100325)
Unit no.
– exclusive number for this Flowmeter. If there are any problems, please refer to this number
Error
(min)
– the number of minutes the device was not measuring because of errors
OK
(min) - the number of minutes that the device measured correctly
Diameter
– this item shows the nominal sensor diameter that is currently configured for the given flowmeter
Firmware No
.
– this shows the current firmware version
Flow Qn
– Nominal flow. Real value = Actual value / 1000
Excitation Counter
– the number of excitations
9.13. Display
Modbus
register
Modbus
address
No. of
bytes
Data
type
No. of
decimal
Min.
Value
Max.
Value
Default Description
Read/
Write
1500
1499
4
Longint
-
0
4
2
Unit Flow (+/-
), 0=UKG/min,
1=USG/min,
2=m3/h,
3=l/min,
4=l/s
R/W
1502
1501
4
Longint
-
0
3
2
Unit Volume,
0=UKG,
1=USG,
2=m3,
3=l
R/W
1504
1503
4
Longint
0
0
100
50
Contrast [%]
R/W
Unit Flow
– actual flow unit (default m
3
/h)
Unit Volume
– totalizer unit (default m
3
)
Contrast
– Set display’s contrast