16
Disinfection, Cleaning and Maintenance
Parts lists and circuit diagrams
These are available on request from ArjoHuntleigh
or its approved distributors.
Spare parts, if required, are available from
ArjoHuntleigh or its approved distributors.
Special tools are required for certain component
replacement.
Periodic Testing
To be carried out at weekly intervals:
For normal operation - raise and lower the resident
support arms using the hand control and dual control
switch, this is to test for full and efficient movement.
Emergency Stop
:
Test the emergency stop facility by operating the
hand control to raise or lower the resident support
arms, and whilst operating, press in the emergency
stop button. (See inset to Fig. 1).
Powered movement should stop immediately.
Reset to normal function by resetting out the button.
Repeat this test using the dual control switch.
Adjustable Width Chassis Function:
Open and close the chassis legs to check for full and
efficient movement.
General Equipment Condition:
A general visual inspection of all external parts
should be carried out, and all functions should be
tested for correct operation, to ensure that no
adverse damage has occurred during use.
Environmental Advice
This device is marked with the WEEE
symbol (crossed-out wheeled bin) to
indicate that it is electronic equipment
covered by the Directive 2002/96/EC on
waste electrical and electronic
equipment. This is a European directive
but applies worldwide. In European
countries the WEEE symbol reminds
you that all electrical and electronic
products must be taken to a separate
collection at the end of their working
life. Do not dispose of this product in
normal domestic or commercial waste -
contact your local authority for advice on
disposal.
Slings
Warning:
If in any doubt about the
correct functioning of the
Sara 3000
,
withdraw it from use and contact
ArjoHuntleigh Service Department.
Warning:
The slings shall be checked
before using with each resident and if
necessary washed according to
instructions on the sling.
With regard to laundering, slings shall
not be classified as linen.
Slings shall be cleaned and disinfected
only in strict accordance with the
manufacturers instructions.
Mechanical pressure shall be avoided
during the washing and drying procedure
e.g. rolling or pressing, as these can
damage parts vital to the safe and
comfortable operation of the sling.
Caution:
Wash using normal
detergents, do not iron. Follow the
instructions on the label of the sling.
It is essential that the sling attachment
cords, the slings, their straps and
attachment clips are carefully inspected
before each and every use -as indicated in
the Preventive Maintenance Schedule.
If the slings, cords or straps are frayed, or
the clips damaged, the sling or attachment
cord should be withdrawn from use
immediately and replaced.