6. Decontamination
The following processes are recommended, but should be adapted to comply with
the local or national guidelines (Decontamination of Medical Devices) which may
apply within the Healthcare Facility or the country of use. If you are uncertain, you
should seek advice from your local Infection Control Specialist.
The
Flowtron Hydroven 3
system should be routinely decontaminated between
patients and at regular intervals while in use; as is good practice for all reusable
medical devices.
To clean
Clean all exposed surfaces and remove any organic
debris by wiping with a cloth moistened with a simple
(neutral) detergent and water.
Do not allow water or cleaning solutions to collect on
the surface of the pump.
Chemical Disinfection
We recommend a chlorine-releasing agent, such as
sodium hypochlorite, at a strength of 1,000ppm
available chlorine (this may vary from 250ppm to
10,000ppm depending on local policy and
contamination status).
Wipe all cleaned surfaces with the solution, then wipe
using a cloth moistened with water and dry thoroughly.
Alcohol based disinfectants (strength 70%) may be used
as an alternative.
Ensure the product is dry before storage.
WARNING
Remove the electrical supply to the pump by disconnecting the
mains power cord from the mains power supply before cleaning.
Protective clothing should always be worn when carrying out
decontamination procedures.
Caution
Do not use Phenol-based solutions or abrasive compounds or pads
during the decontamination process as these will damage the
surface coating. Avoid immersing electrical parts in water during
the cleaning process. Do not spray cleaning solutions directly onto
the pump. Do not immerse the tubeset in water.