Caregiver Obligations
Caregiver obligations shall be carried out by personnel with
sufficient
Carendo
knowledge following the instructions in this
IFU
.
Every day
•
Disinfect.
The
Carendo
shower and toilet chair has to be
disinfected immediately after every use. ArjoHuntleigh
disinfectants and shampoos should be used in
recommended concentrations
. See section
Cleaning and
Disinfection Instructions on page 26.
Every week
•
Visually check the safety belt
. Check the complete length
for fraying, cuts, crack, tears, that no inner fabric is showing
and that no other damage is visible. If the safety belt is found
damaged, replace it.
(See Fig. 1)
•
Visually check all exposed parts
, especially where
personal contact is made by either the patient or caregiver.
Make sure no cracks or sharp edges have developed that
could cause the patient or user injury or that has become
unhygienic. Replace damaged parts.
•
Visually check mechanical attachments
: Visually check
that all screws and nuts are tightened and that there are no
gaps - contact your local ArjoHuntleigh representative if
defects are found.
(See Fig.2)
•
Check hand control and cable
: Visually check the
condition of the hand control and its cable if it is cut or
damaged. Also check the emergency stop, see product
functions. Contact your local ArjoHuntleigh representative
if defects are found.
(See Fig. 3)
Fig. 1
Fig. 2
Fig. 3
Содержание Carendo
Страница 1: ...Carendo with people in mind 04 CC 01_23GB 12 2016 Instructions for Use ...
Страница 9: ...9 6 13 5 12 2 3 3 1 9 14 17 16 15 19 20 24 22 23 8 7 21 17 10 18 11 4 3 ...
Страница 19: ...19 Intentionally left blank ...
Страница 39: ...39 Data Label Attention Symbol Approval Label Identification Label Emergency Lowering Label Battery Label ...