Transport mode (mattresses only)
2
1
Activate:
Push the button down and pull the tubeset connector.
The mattress supports the patient for up to 12 hours.
Resume normal operation:
Reconnect the tubeset.
CPR control (mattresses only)
2
1
Activate:
Press the two CPR release buttons
simultaneously.
Reset:
Push the front of the CPR unit until it clicks
into place.
“Click”
Cleaning and disinfection
Clean and disinfect between patients and at regular
intervals while in use.
WARNING
Remove the electrical supply to the pump before
cleaning by disconnecting the mains power cord
from the mains power supply or removing the
battery pack.
Protective clothing should always be worn when
carrying out decontamination.
Cleaning
1. Clean all exposed surfaces and remove any debris by
wiping with a cloth moistened with a neutral detergent
and water.
2. Dry thoroughly. Do not allow water or cleaning
solutions to collect on the surface of the pump.
Chemical disinfection
Use a chlorine releasing agent, such as sodium
hypochlorite, at a strength of 1000 ppm available chlorine
(from 250 ppm to 10 000 ppm depending on local policy
and contamination status).
Alcohol based disinfectants (strength 70%) may be used
as an alternative.
If an alternative disinfectant is selected, confirm suitability
for use with the chemical supplier prior to use.
1. Wipe all cleaned surfaces with the solution.
2. Wipe with a cloth moistened in water.
3.
Dry thoroughly.
Thermal disinfection
For laundering of the mattress top cover, see label.
Autofirm mode
Activate:
Press and hold
the button.
Comfort indicators flash
and then illuminates in
sequence.
Extend:
Comfort indicators
flash + series of tones:
Press either comfort key
to extend 5 minutes.
Maximum 30 minutes.
1
2
Deactivate:
Press and
hold the button.
2s
Pressure reached:
•
Comfort indicators
illuminated
•
Pump beeps 3 times.
•
Reactive (CLP) indicator
illuminated
2s