13

Troubleshooting

If the equipment fails to operate correctly, the table below suggests 
some simple checks and corrective actions. If these steps fail to 
resolve the problem, contact ArjoHuntleigh or an approved service 
agent.

Product lifetime

The lifetime of this equipment is typically ten (10) years from 
manufacture date. “Lifetime” is defined as the period during which 
the product will maintain the specified performance and safety, 
provided it has been maintained and operated in conditions of 
normal use in accordance with the requirements in these 
instructions.

Symptom

Possible Cause

Action

None of the actuators 
work

Mains cable disconnected

Check the mains cable is 
plugged in and the 
electricity supply is okay

Fuse blown

Check fuse in mains plug 
(where fitted)

Control handset 
disconnected

Check handset is plugged 
into the control box

Functions disabled

Enable the functions on 
handset using the plastic 
key

One actuator does not 
work

Actuator cable 
disconnected

Check actuator cable is 
plugged into both the 
actuator and control box

Function disabled

Enable a function on 
handset using the plastic 
key

Castor does not rotate or 
swivel

Brake applied

Release brake

Содержание 2132B

Страница 1: ...Akron Streamline Europe Continental with people in mind 792 692 07 10 2017 Instructions for Use ...

Страница 2: ... one of continuous improvement we reserve the right to modify designs without prior notice The content of this publication may not be copied either wholly or in part without the consent of ArjoHuntleigh WARNING To avoid injury always read this Instructions for Use and accompanied documents before using the product Mandatory to read the Instructions for Use ...

Страница 3: ...e 7 Breathing Hole and Plug 8 Head and Foot Sections 8 A and B Type Head Section Operating Lever 8 C Type Head Section Operating Lever 8 Foot Section Operating Lever 9 C Type Head Section Arm Rests 9 5 Product Care 10 Cleaning Decontamination 10 To Clean 10 Chemical Disinfection 10 Preventive maintenance 11 Troubleshooting 13 Product lifetime 13 6 Accessories 14 Detachable Head Support AHC1 15 Eur...

Страница 4: ... later on Read and understand these instructions before operating the couch Caregivers must be trained in the proper use of this product its functions and controls and any accessories These instructions are mandatory for the safe and effective use of this product including the safety of patients and caregivers Unauthorised modifications or repairs to this product may affect its safety and will inv...

Страница 5: ...quipment should be verified by observation Medical Electrical Equipment maybe affected by portable and or mobile RF communications equipment e g mobile telephones These products are not intended for Patient Transportation Maximum Load 225kg 495lbs must be distributed evenly Do not concentrate weight on either side of the couch Plug acts as a disconnecting device Do not allow children to play with ...

Страница 6: ...uirements of applicable EMC standards The use of accessories not specified by the manufacturer may result in increased emissions by or decreased immunity of the equipment affecting its performance Do not exceed the specified duty cycle as stated on the Product Identification Label ...

Страница 7: ...itional accessories also available However these products are not designed to be used as a bed and are not suitable for sleeping on You must read and fully understand this manual before using the system This manual is your introduction to the following STREAMLINE couches Streamline Europa Continental model 2132B without drainage Streamline Europa Continental model 2132C without drainage Streamline...

Страница 8: ...2 Product overview 1 Upholstery with breathing hole 2 Gas strut lever 3 Castors 4 Lifting arms 5 Rubber pedals 6 Electric actuator 7 Paper roll holder 1 7 4 3 5 2 6 ...

Страница 9: ...tions This couch is used to position patients for face down spine treatment for physiotherapy treatment Contraindications The safe working load SWL of the couch 225kg The maximum recommended patient weigh 210kg Accessories including attached loads 15kg WARNING To ensure the patient can use the couch safely their age size and condition should be assessed by a clinically qualified person ...

Страница 10: ...lug is damaged the complete assembly must be replaced by authorised service personnel Do not remove the fitted plug or use a rewireable plug or adapter Do not allow the power supply cord to trail on the floor where it may cause a trip hazard Ensure during couch use no obstacles are in its path Make sure the power supply cord does not become entangled with moving parts of the couch Damaged cables c...

Страница 11: ...5 Caution Do NOT insert hands or feet into the mechanism as this could result in serious injury ...

Страница 12: ... position B To enable a function fit the plastic key into the two holes in the lockout switch and turn the switch anti clockwise to the Unlocked position C WARNING Safely position patient before elevating couch Position all electrical cables away from the moving parts Position hand foot switch where accidental operation is not possible Lockout handset Press to move the section up Press to move the...

Страница 13: ...switch FBS Lift tube up to lower couch Press tube down to raise couch WARNING Lock all couch wheels castors prior to patient transfer or treatment These couches are not designed for the transportation of patients Disconnect and coil electrical cord away from moving parts prior to repositioning couch Castors locked Castors unlocked ...

Страница 14: ...ution Support upholstery section while manually altering the angle Apply pressure to ensure the upholstery is locked If movement is detected seek advice A Press lever UP to alter section angle B Support section when altering angle C Release lever to lock section in position A Press lever UP to alter section angle B Support section when altering angle C Release lever to lock section in position ...

Страница 15: ...ction Arm Rests A Press lever UP to alter section angle B Support section when altering angle C Release lever to lock section in position A Tilt Armrest towards head section B Move armrest into position C Release lever to lock section in position ...

Страница 16: ...trength 70 may be used as an alternative Ensure the product is dry before storage If an alternative disinfectant is selected from the wide variety available we recommend that suitability for use is confirmed with the chemical supplier prior to use WARNING Do not use Alkalis or other organic solvents for routine cleaning Replace torn or damaged vinyl immediately Always cover vinyl with fresh paper ...

Страница 17: ...onments or where required by local regulations Failure to carry out these checks or continuing to use the product if a fault is found may compromise the safety of both the patient and caregiver Preventive maintenance can help to prevent accidents To isolate the couch from the electrical supply disconnect the electrical power cable from socket outlet Product cannot be maintained and serviced while ...

Страница 18: ...al backrest footrest Check operation of the castors Examine the power supply cord and mains plug if damaged replace the complete assembly do not use a rewireable plug Examine all accessible flexible cables for damage and deterioration Check all accessible nuts bolts and other fasteners are present and correctly tightened Check any accessories fitted to the couch paying particular attention to fast...

Страница 19: ...itions of normal use in accordance with the requirements in these instructions Symptom Possible Cause Action None of the actuators work Mains cable disconnected Check the mains cable is plugged in and the electricity supply is okay Fuse blown Check fuse in mains plug where fitted Control handset disconnected Check handset is plugged into the control box Functions disabled Enable the functions on h...

Страница 20: ...l Accessory Accessory description Accessory number EC Weight of accessory kg Streamline Europa Continental model 2132B without drainage Streamline Europa Continental model 2132C without drainage Streamline Europa Continental model 2162C split leg with drainage Detachable Head Support AHC1 0 6 Europa Paper Roll Holder A8612 0 5 ...

Страница 21: ...ient by A Unfastening Velcro strap B Position where required C Fasten Velcro strap to secure Replace paper roll by A Gently spring bracket out B Insert paper roll rod between brackets C Load paper roll and release brackets to secure Caution Ensure all accessories are fully tightened prior to use ...

Страница 22: ...nc with respect to electrical shock fire mechanical hazards and other specified hazards only in accordance with the Approvals and List of Standards Environmental Protection This product is marked with the crossed out wheelie bin symbol to indicate that it is covered by EU Directive 2002 96 EC on Waste Electrical and Electronic Equipment WEEE This symbol indicates that all electrical and electronic...

Страница 23: ... Handle with care Do not drop Avoid shock or violent impact This equipment should be stored in a clean dry and well ventilated area which meets the following conditions Temperature 10 C to 50 C Relative humidity 20 to 90 non condensing Atmospheric pressure 700 to 1060hPa ...

Страница 24: ...and Upholsteries Model number Waste Electrical and Electronic Equipment WEEE do not dispose of this product in general household or commercial waste Functional earth Does not provide any protection against electrical shock Used due to the EMC requirements Caution Electromagnetic compatibility EMC This product complies with the requirements of applicable EMC Standards The use of accessories not spe...

Страница 25: ...ed is intended for use in the electromagnetic environment specified below The customer or the user of the bed should assure that it is used in such an environment Emissions test Compliance Electromagnetic environment guidance RF emissions CISPR 11 Group 1 The couch uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in ...

Страница 26: ...line 1kV 1kV Mains power quality should be that of a typical commercial or hospital environment 2kV line to earth 2kV 2kV Voltage dips short interruptions and voltage variations on power supply input lines IEC 61000 4 11 2004 70 UT 30 dip in UT for 25 cycles 40 UT 60 dip in UT for 5 cycles 5 UT 95 dip in UT for 0 5 cycle 5 UT 95 dip in UT for 5 sec 70 of input voltage for 500ms 40 of input voltage...

Страница 27: ...electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80MHz and 800MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and peo...

Страница 28: ... that the product has been cleaned to the outside of the package 4 Mark the package Service Department Manufactured in Poland by ArjoHuntleigh AB As part of the ongoing development programme this company reserves the right to modify specifications and materials of this couch without notice A full technical description of the Couch can be found in the service documentation please contact ArjoHuntle...

Страница 29: ...Intentionally left blank ...

Страница 30: ...Intentionally left blank ...

Страница 31: ...logy Research Park P O Box 214742 Dubai United Arab Emirates Tel 971 0 4 447 0942 E mail Info ME getinge com NEDERLAND ArjoHuntleigh Nederland BV Biezenwei 21 4004 MB TIEL Postbus 6116 4000 HC TIEL Tel 31 0 344 64 08 00 Fax 31 0 344 64 08 85 E mail info nl arjohuntleigh com NEW ZEALAND ArjoHuntleigh Ltd 41 Vestey Drive Mount Wellington NZ AUCKLAND 1060 Tel 64 0 9 573 5344 Free Call 0800 000 151 Fa...

Страница 32: ...obilisation safe patient handling venous thromboembolism prevention pressure injury prevention hygiene routines bariatric care and diagnostics With extensive know ledge and experience we strive to improve efficiency and ensure a safer and dignified environment for caregivers and their patients ArjoHuntleigh AB Hans Michelsensgatan 10 211 20 Malmö Sweden www arjohuntleigh com BARCODE including huma...

Отзывы: