background image

The sensor head must remain 
untouched during this time

After approximately 4 seconds the 

Install Sensor Screen

 will display

1

Remove the Provizio SEM Scanner single-use sensor 
from its packaging and place onto the sensor head. 
You will hear a click when the sensor is placed 
correctly

Each patient scanning session requires a new sensor

2

3

Press the 

Next Screen Button

 to continue, the 

scanner will test the installation of the sensor 
and then display the Mode of Operation Screen

0

3

1

4

2

5

6

7

4  PROVIZIO SEM SCANNER • PRODUCT TRAINING

Start-up

By removing from the 

Charging Hub 

or
By pressing the 

Action Button

 

until the screen illuminates and 
the Splash Screen displays 

Turn on Provizio SEM Scanner

Installing the Sensor

Содержание Provizio SEM Scanner

Страница 1: ...Product training Provizio SEM Scanner PRESSURE INJURY PREVENTION...

Страница 2: ...Summary Before you start Start up Mode of operation Scan Reset Clean Data upload Charge power down 0 1 2 3 4 5 6 7 2 PROVIZIO SEM SCANNER PRODUCT TRAINING...

Страница 3: ...rt Action button On device Patient ID Sensor Patient barcode scanner Available in Provizio V2 Status indicator SEM delta Charging hub 0 3 1 4 2 5 6 7 Battery gauge SEM value 3 PROVIZIO SEM SCANNER PRO...

Страница 4: ...or is placed correctly Each patient scanning session requires a new sensor 2 3 Press the Next Screen Button to continue the scanner will test the installation of the sensor and then display the Mode o...

Страница 5: ...the Barcode Button on the Barcode Scanning Screen Position the scanner and use the green light to aim the barcode reader at the patient barcode Once the scanner has identified the barcode it will emi...

Страница 6: ...Gateway Dashboard installed on your WiFi network Option C Manual Patient ID Button lower right corner Pressing the Manual Patient ID Button in the lower right corner will allow you to manually enter t...

Страница 7: ...will flash blue and emit a short audio tone Lift the sensor off the skin the session circles under the body location will display completely filled in when the measurement is complete Repeat above ste...

Страница 8: ...ng 1 Reading 4 taken approximately 1 5 cm right of center of Reading 1 Readings 5 and 6 is taken outside the sacral bone area Reading 5 Taken approximately 4 5 cm left of center of Reading 1 Reading 6...

Страница 9: ...e side as the little toe fifth toe Avoid placement of the Provizio SEM Scanner on the lateral malleolus Medial inner side of heels Take one 1 Provizio SEM Scanner reading on the medial side of the ass...

Страница 10: ...ll readings taken in encrypted form until uploaded on the Gateway Dashboard If using the Manual Charting Option the SEM readings are deleted once the scanning session is left Remove the sensor by gent...

Страница 11: ...no instructions are provided the recommended disposal of used wipe s is medical waste do not flush down the toilet ENGLISH OUS indications Page 36 of 50 OTH SEM IFU OUS 0359 Rev G Scanner Disinfection...

Страница 12: ...e Charging Hub the scanner initiates wireless communications and connects to the Gateway Dashboard All stored encrypted data sessions are uploaded to the Gateway Dashboard and deleted from the scanner...

Страница 13: ...sabled The display shows the status of the data upload and battery charge level The scanner is fully charged when five green bars are illuminated To power down the scanner can be turned off by pressin...

Страница 14: ...ed to promote a safe and dignified experience through patient handling medical beds personal hygiene disinfection diagnostics and the prevention of pressure injuries and venous thromboembolism With ov...

Отзывы: