background image

Clean and disinfect

Clean and disinfect the Provizio SEM Scanner following  
the cleaning and disinfection procedure in the User Manual.

Install a new sensor if starting another scanning session.

0

3

1

4

2

5

6

7

 

 

ENGLISH OUS indications 

 

Page 1 of 50 

OTH-SEM-IFU-OUS-0359 Rev G 

 

 

 
 

 

Provizio

®

 SEM Scanner S 

Provizio

®

 SEM Scanner FH 

 

User Manual 

 

 

 

 

 

ENGLISH OUS indications 

 

Page 35 of 50 

OTH-SEM-IFU-OUS-0359 Rev G 

 

Scanner Cleaning Instructions 

The Provizio SEM Scanner should be cleaned using nonabrasive cleaning wipes and wetted with 

agents such as isopropyl alcohol or mild detergents. 

(1)  Using a fresh wipe, verify the wipe is moist.   

(2)  Thoroughly wipe all surfaces of the Provizio SEM Scanner for 45 seconds to remove dirt 

and debris from the SEM Scanner as shown in Figure 26 and Figure 27.  Continue wiping 

for the entire 45 seconds even if there is no visible dirt and debris.   Do not reuse the 

wipes. 

 

 

Figure 24 Clean the Sides and Back  

of the Scanner 

 

Figure 25 Clean the Front of the Scanner 

 

(3)  Dispense  a  fresh  towelette  wipe  from  the  dispenser  and  verify  the  wipe  is  moist.  

Thoroughly wipe all surfaces of the Scanner for a minimum of 30 seconds to remove any 

remaining visible soil. Continue wiping for the entire 30 seconds even if there is no visible 

dirt and debris. The device surface should be fully coated with the wipe solution.  Do not 

reuse the wipes. 

(4)  Dispense a fresh towelette wipe from the dispenser and verify it is moist.  Thoroughly 

wipe all surfaces of the Provizio SEM Scanner for a minimum of 30 seconds. Continue 

wiping for the entire 30 seconds even if there is no visible dirt and debris.   The device 

surface should be fully coated with the wipe solution.  Do not reuse the wipes. 

(5)  Dispose the used wipe(s) as instructed by the wipe manufacturer.  If no instructions are 

provided, the recommended disposal of used wipe(s) is medical waste; do not flush down 

the toilet.   

 

 

ENGLISH OUS indications 

 

Page 36 of 50 

OTH-SEM-IFU-OUS-0359 Rev G 

 

Scanner Disinfection Instructions 

To properly disinfect the external surfaces of the Provizio SEM Scanner using CaviWipe™ by 

Metrex  Research,  follow  the steps  below.    Wearing  gloves  and an  apron  during  cleaning  and 

disinfection of the device is recommended.  Please read the CaviWipe™ label instructions for a 

full list of precautions and use guidelines: manufacturers are periodically updated.  
If other wipes are preferred, select nonabrasive intermediate-level cleaning and disinfection wipes. 

Users must follow the wipe manufacturer’s cleaning and disinfection instructions.  The selected 

wipe should not contain any of the chemical agents listed in the Warranty and Disclaimer section.   

7.3.1

 

Disinfection Steps using CaviWipe 

Note:    When  wiping  the  Provizio  SEM  Scanner  with  the  wipe,  you  are  applying  the  cleaning  and 

disinfecting agent onto the scanner. 
Before starting the disinfection, first clean the scanner as described above. 

(1)  Dispense  fresh CaviWipe™ towelette wipe from the dispenser and verify the wipe is 

moist.   

(2)  Thoroughly wipe all surfaces of the scanner for 45 seconds to remove dirt and debris 

from the scanner as shown in Figure 26 and Figure 27.  Continue wiping for the entire 45 

seconds even if there is no visible dirt and debris.   Do not reuse the wipes.  

 

Figure 26 Disinfect the Sides and Back  

of the Scanner 

 

Figure 27 Disinfect the Front of the Scanner 

 

(3)  Dispense  a  fresh  towelette  wipe  from  the  dispenser  and  verify  the  wipe  is  moist.  

Thoroughly wipe all surfaces of the scanner for a minimum of 30 seconds to remove any 

remaining visible soil. Continue wiping for the entire 30 seconds even if there is no visible 

dirt and debris. The scanner surface should be fully coated with the wipe solution.   Do 

not reuse the wipes. 

11  PROVIZIO SEM SCANNER • PRODUCT TRAINING

Содержание Provizio SEM Scanner

Страница 1: ...Product training Provizio SEM Scanner PRESSURE INJURY PREVENTION...

Страница 2: ...Summary Before you start Start up Mode of operation Scan Reset Clean Data upload Charge power down 0 1 2 3 4 5 6 7 2 PROVIZIO SEM SCANNER PRODUCT TRAINING...

Страница 3: ...rt Action button On device Patient ID Sensor Patient barcode scanner Available in Provizio V2 Status indicator SEM delta Charging hub 0 3 1 4 2 5 6 7 Battery gauge SEM value 3 PROVIZIO SEM SCANNER PRO...

Страница 4: ...or is placed correctly Each patient scanning session requires a new sensor 2 3 Press the Next Screen Button to continue the scanner will test the installation of the sensor and then display the Mode o...

Страница 5: ...the Barcode Button on the Barcode Scanning Screen Position the scanner and use the green light to aim the barcode reader at the patient barcode Once the scanner has identified the barcode it will emi...

Страница 6: ...Gateway Dashboard installed on your WiFi network Option C Manual Patient ID Button lower right corner Pressing the Manual Patient ID Button in the lower right corner will allow you to manually enter t...

Страница 7: ...will flash blue and emit a short audio tone Lift the sensor off the skin the session circles under the body location will display completely filled in when the measurement is complete Repeat above ste...

Страница 8: ...ng 1 Reading 4 taken approximately 1 5 cm right of center of Reading 1 Readings 5 and 6 is taken outside the sacral bone area Reading 5 Taken approximately 4 5 cm left of center of Reading 1 Reading 6...

Страница 9: ...e side as the little toe fifth toe Avoid placement of the Provizio SEM Scanner on the lateral malleolus Medial inner side of heels Take one 1 Provizio SEM Scanner reading on the medial side of the ass...

Страница 10: ...ll readings taken in encrypted form until uploaded on the Gateway Dashboard If using the Manual Charting Option the SEM readings are deleted once the scanning session is left Remove the sensor by gent...

Страница 11: ...no instructions are provided the recommended disposal of used wipe s is medical waste do not flush down the toilet ENGLISH OUS indications Page 36 of 50 OTH SEM IFU OUS 0359 Rev G Scanner Disinfection...

Страница 12: ...e Charging Hub the scanner initiates wireless communications and connects to the Gateway Dashboard All stored encrypted data sessions are uploaded to the Gateway Dashboard and deleted from the scanner...

Страница 13: ...sabled The display shows the status of the data upload and battery charge level The scanner is fully charged when five green bars are illuminated To power down the scanner can be turned off by pressin...

Страница 14: ...ed to promote a safe and dignified experience through patient handling medical beds personal hygiene disinfection diagnostics and the prevention of pressure injuries and venous thromboembolism With ov...

Отзывы: