Clean and disinfect
Clean and disinfect the Provizio SEM Scanner following
the cleaning and disinfection procedure in the User Manual.
Install a new sensor if starting another scanning session.
0
3
1
4
2
5
6
7
ENGLISH OUS indications
Page 1 of 50
OTH-SEM-IFU-OUS-0359 Rev G
Provizio
®
SEM Scanner S
Provizio
®
SEM Scanner FH
User Manual
ENGLISH OUS indications
Page 35 of 50
OTH-SEM-IFU-OUS-0359 Rev G
Scanner Cleaning Instructions
The Provizio SEM Scanner should be cleaned using nonabrasive cleaning wipes and wetted with
agents such as isopropyl alcohol or mild detergents.
(1) Using a fresh wipe, verify the wipe is moist.
(2) Thoroughly wipe all surfaces of the Provizio SEM Scanner for 45 seconds to remove dirt
and debris from the SEM Scanner as shown in Figure 26 and Figure 27. Continue wiping
for the entire 45 seconds even if there is no visible dirt and debris. Do not reuse the
wipes.
Figure 24 Clean the Sides and Back
of the Scanner
Figure 25 Clean the Front of the Scanner
(3) Dispense a fresh towelette wipe from the dispenser and verify the wipe is moist.
Thoroughly wipe all surfaces of the Scanner for a minimum of 30 seconds to remove any
remaining visible soil. Continue wiping for the entire 30 seconds even if there is no visible
dirt and debris. The device surface should be fully coated with the wipe solution. Do not
reuse the wipes.
(4) Dispense a fresh towelette wipe from the dispenser and verify it is moist. Thoroughly
wipe all surfaces of the Provizio SEM Scanner for a minimum of 30 seconds. Continue
wiping for the entire 30 seconds even if there is no visible dirt and debris. The device
surface should be fully coated with the wipe solution. Do not reuse the wipes.
(5) Dispose the used wipe(s) as instructed by the wipe manufacturer. If no instructions are
provided, the recommended disposal of used wipe(s) is medical waste; do not flush down
the toilet.
ENGLISH OUS indications
Page 36 of 50
OTH-SEM-IFU-OUS-0359 Rev G
Scanner Disinfection Instructions
To properly disinfect the external surfaces of the Provizio SEM Scanner using CaviWipe™ by
Metrex Research, follow the steps below. Wearing gloves and an apron during cleaning and
disinfection of the device is recommended. Please read the CaviWipe™ label instructions for a
full list of precautions and use guidelines: manufacturers are periodically updated.
If other wipes are preferred, select nonabrasive intermediate-level cleaning and disinfection wipes.
Users must follow the wipe manufacturer’s cleaning and disinfection instructions. The selected
wipe should not contain any of the chemical agents listed in the Warranty and Disclaimer section.
7.3.1
Disinfection Steps using CaviWipe
Note: When wiping the Provizio SEM Scanner with the wipe, you are applying the cleaning and
disinfecting agent onto the scanner.
Before starting the disinfection, first clean the scanner as described above.
(1) Dispense fresh CaviWipe™ towelette wipe from the dispenser and verify the wipe is
moist.
(2) Thoroughly wipe all surfaces of the scanner for 45 seconds to remove dirt and debris
from the scanner as shown in Figure 26 and Figure 27. Continue wiping for the entire 45
seconds even if there is no visible dirt and debris. Do not reuse the wipes.
Figure 26 Disinfect the Sides and Back
of the Scanner
Figure 27 Disinfect the Front of the Scanner
(3) Dispense a fresh towelette wipe from the dispenser and verify the wipe is moist.
Thoroughly wipe all surfaces of the scanner for a minimum of 30 seconds to remove any
remaining visible soil. Continue wiping for the entire 30 seconds even if there is no visible
dirt and debris. The scanner surface should be fully coated with the wipe solution. Do
not reuse the wipes.
11 PROVIZIO SEM SCANNER • PRODUCT TRAINING