Risk
Definitions
Procedure
Low
Normal use or low risk situations include
patients having intact skin and no known
infection and the probes have not been
contaminated with blood.
1. Remove soiling, wipe with a mild
neutral detergent and then wipe
with a cloth dampened in water.
2. Completely dry with a clean lint
free cloth.
Medium
The patient has a known infection, skin
is not intact, the part is heavily soiled.
1. Follow low risk procedure then
wipe with a cloth dampened in
Sodium Hypochlorite (1,000ppm).
2. After two minutes wipe with a cloth
dampened in water and then dry
with a clean lint free cloth.
High
This procedure should only be used
when the part has been contaminated
by blood.
1. Follow low risk procedure then
wipe with a cloth dampened in
Sodium Hypochlorite (10,000ppm).
2. After two minutes wipe with a cloth
dampened in water and then dry
with a clean lint free cloth.
WARNING
Repeated and unnecessary use of concentrated solutions will result in damage to the product. Do not
allow Sodium Hypochlorite solutions to come into contact with metal parts.
The use of disinfectant materials, other than those listed, is the responsibility of the user for their efficacy and
compatibility with the device.
Approved Products
DO NOT USE
LINGET ANIOS
SURFA'SAVE
WIP'ANIOS Excel
Surfanios Premium
Clinell Alcohol Wipes (Red)
Clinell Alcohol Wipes Plus+ (Red)
Clinell Universal Wipes (Green)
Clinell Universal Spray (Green)
Clinell Detergent Range (Yellow)
WARNING
After using chemicals ALWAYS rinse off / remove the chemical with absorbent material, dampened in
clean water and dry with a clean cloth.
12.4 NIBP Cuff
For cleaning and disinfection, refer to instructions supplied by the manufacturer.
12.5 Transducer Belts
Re-useable belts can be hand-washed at 40°C maximum using a washing powder or mild detergent solution
following the detergent manufacturer’s guidance. Rinse with clean water and dry without using heat.