Mattress and Seat Systems Service Manual
Repair
Page 40
7
Installing a Cell
Note:
There are several different cell assemblies in the various mattress replacements, overlays and
seat systems detailed in this service manual. Make sure the cell assembly is the correct type for
the particular mattress or seat, and also for the position within the mattress (where applicable).
Note:
Make sure the orientation of the cell is correct (Refer to Figure 25). The seams should be facing
the feet of the mattress.
Note:
Installing the T or elbow connectors into the manifold tubes or side-entry grommets on the cells
is made easier if the parts are wetted with methylated spirit immediately prior to insertion.
All Mattress Replacement and Overlay Systems except Alpha Trancell II
7.1
Feed the cell through the loops in the loop base sheet assembly, with the cell T-connector
adjacent to the manifold assembly.
7.2
Carefully insert the two ends of the T-connector into the manifold tubes.
7.3
Connect the press studs that secure each end of the cell to the loop base sheet assembly.
7.4
Install the top cover (Refer to Page 37, Section 4).
Alpha Trancell II Mattress Overlay only
7.5
To install the mattress overlay assembly, Refer to Page 44, Section 12.
All Seat Systems
7.6
Feed the cell through the loops in the loop base sheet assembly, with the side-entry grommet on
the end of the cell adjacent to the T-connector on the manifold assembly.
7.7
Carefully insert the end of the T-connector (or elbow connector on the end cells) into the side-
entry grommet on the cell.
7.8
Connect the press studs that secure each end of the cell to the loop base sheet assembly.
7.9
Install the top cover (Refer to Page 37, Section 4).
Figure 25 - Cell Orientation
Cell
Manifold Tubes
T-Connector
Side-Entry
Grommet
Cell Seam
Feet End
Cell Loop