10
• Connect the
Slop Hopper
to protective grounding
and to the supply voltage stated on the type plate.
+
Check that the connection has suitable
overcurrent protection. The correct fuse
rating is stated on the type plate.
• All cables must be classified for at least 90°C.
8. Perform the following function check on the
Slop
Hopper
after installation.
• Check that the
Slop Hopper
is connected to the
correct supply voltage and that it is protected by
a fuse of the correct rating as shown on the type
plate.
• Check that the temperature of the incoming
mains water is correct. The temperature must be
in the range 3 - 25°C.
• Check that there are no water leaks. Tighten
connections and pipe couplings.