Ranger
Blood/Fluid Warming System
10
En
glish
Ranger
Blood/Fluid Warming System
En
glish
11
Specifications
Physical characteristics
WARMING
UNIT
4.5 in. (11 cm) high x 7.5 in. (19 cm) wide x 10 in. (25 cm) long;
wt.: 7 lb. 7 oz. (3.4 kg)
CLASSIFICATION
Classified under EN 60601-1 Guidelines (and other national versions of the Guidelines) as
Class I, Type B, Ordinary Equipment. Not suitable for use in the presence of flammable
anesthetic mixtures with air or with oxygen or nitrous oxide. Classified by Underwriters
Laboratories Inc. with respect to electric shock, fire, and mechanical hazards only, in accordance
with UL 2601-1 and in accordance with Canadian/CSA C22.2 No. 601.1.
Electrical characteristics
INPUT
VOLTAGE
100-120 or 220-240 VAC
OPERATING
FREQUENCY
100-120 VAC, 50/60 Hz
220-240 VAC, 50 Hz
MAXIMUM
HEATING
POWER
900 W
FUSE
10 Amp
Leakage current
Meets leakage current requirements in accordance with UL 2601-1 and EN 60601-1.
Environmental conditions
OPERATING
TEMPERATURE
RANGE
15° to 40°C (59° to 104°F)
STORAGE
TEMPERATURE
RANGE
-20° to 45°C (-4° to 113°F)
OPERATING
HUMIDITY
10 to 85% RH, noncondensing
ATMOSPHERIC
PRESSURE
RANGE
50 kPa to 106 kPa
Temperature characteristics
SET
POINT
TEMPERATURE
41°C
OVER
-
TEMPERATURE
ALARM
43°C
UNDER
-
TEMPERATURE
ALARM
33°C
OVER
-
TEMPERATURE
CUTOFF
46°C