HEALTH AND SAFETY
Installation and service of the equipment should
be performed by authorized and trained personnel.
Safety measures and local guidelines must be
followed. Use good practices wearing personal
protective equipment (PPE) during the installation
and service of the equipment.
Electricity – Aries Filterworks lab water equipment
should be powered by a grounded GFI electrical
connection. Do not open the system unless the
system is unplugged. Before any installation or service
is performed on the equipment, make sure the power
is isolated. To unplug the system, do not pull directly
on the cord but grip the plug and remove it from
the outlet.
Ultraviolet Light – Aries Filterworks lab water systems
may include an ultraviolet light lamp in a stainless-steel
housing. Make sure all power is removed prior to
accessing the side panel to perform any UV
maintenance. Make sure gloves are used when
handling the UV bulb. (Gloves must be worn when
handling the UV bulb. Which protects the integrity of
the bulb) Do not look directly at the UV lamp while the
system is operating.
Water Supply – Good piping practices should be
followed feeding water to the lab equipment. Isolation
valves, pressure gauges and bleed valves are recom-
mended on the main water supply. Pressure should
be regulated to meet the minimum and maximum
ZWLJPÄJH[PVUZZ[H[LKMVY[OLLX\PWTLU[
Sanitization – Make sure PPE is worn during
any sanitization process. The Aries system uses
chlorine as a disinfectant. Chlorinated water will be
KPZJOHYNLKK\YPUNZHTWSPUNHUKÅ\ZOPUNZLX\LUJLZ
Avoid breathing fumes and ensure the area is
Z\ѝJPLU[S`]LU[PSH[LK
SYSTEM OVERVIEW
;OL(YPLZ3HI^H[LYZ`Z[LTKLSP]LYZ;`WL4£
X\HSP[`^H[LYVUKLTHUK>H[LYPZW\YPÄLKPUHZ[HNLK
process consisting of high-purity ion exchange resins
to remove dissolved minerals and internal recirculation
[VTHPU[HPUW\YP[`([KPZJOHYNLHÄUHSTÄS[LY
removes particulates and bacteria to attain CLSI /
(:;4;`WL^H[LYZWLJPÄJH[PVUZ(KKP[PVUHSVW[PVUHS
technologies are incorporated depending on model.
9LZPZ[P]P[`TVUP[VYWYV]PKLZMLLKIHJRVM[OLÄUHS
water quality.
The Aries system is designed to recirculate through
multiple technologies to maintain purity. A timer
controller runs system automatically for 30 minutes
every 2 hours to maintain the integrity of the system.
As the system transitions from “Operating” mode
to “Stand By” mode, the recirculation pump,
solenoid, resistivity display and UV (if installed) will be
H\[VTH[PJHSS`[\YULKVќ]PH[OL[PTLY(>(2,I\[[VU
on the side of the system will reset the cycle timer
back to “Operating” condition.
Options available with the Aries lab system are UV,
recirculating dispensing gun, secondary direct feed
port, and wall mounting bracket.