Argo Lab AM 20-D Скачать руководство пользователя страница 3

 

UM Agitatore meccanico digitale ad asta AM 20-D ENG rev.0 12.05.2020 

Display 

 

 

 

 

 

Switching on the instrument 

 

Place the agitator on a stable surface and connect the power cable. 

 

Switch on the instrument. 

 

The instrument starts a short self-diagnosis process. 

 

When the initialisation is finished, the "set" icon is displayed and at the same time the value setting area flashes 
to indicate the possibility of setting the speed value. 

 

Turn the SPEED knob to set the stirring speed. 

 

Pressing the SPEED knob the LCD display stops flashing and the instrument starts to shake. 

 

Press the speed knob again, the LCD display flashes and the parameters can be changed again. 

 

 

 

Overload protection 

The rod agitator operates continuously, therefore in the event that the liquid to be stirred becomes too viscous, i.e. an 
overload occurs, the motor power is automatically regulated electronically. When the "torque limit" value is reached, 
the  overload  protection  function  is  started  and  at  the  same  time  the  red  LED  starts  flashing  to  indicate  a  small 
overload.La protezione contro il sovraccarico interviene nei seguenti casi: 

 

When the value of the set speed is not suitable for the current average viscosity; 

 

When the crankshaft is braked / blocked. 
 
 

Malfunctions

 

 

The instrument does not start when the stirring function is started. 

o

 

Check that the power cable is connected. 

 

The speed does not reach the set value 

o

 

The set speed value is not sufficient for the viscosity level of the stirred liquid 

 

The agitation suddenly stops 

o

 

The overload protection light turns red, the display shows "Er 03", indicating that the current fault is 
"overload protection". 

o

 

The overload protection light turns red, the display area shows "Er 04", indicating that the current 
fault is "motor protection". 
 
 
 
 
 
 

Set/actual value 

Содержание AM 20-D

Страница 1: ... also be firmly connected to the device The instrument can be disconnected by disconnecting it from the power supply or by disconnecting the cable The operating voltage indicated on the instrument label must correspond to that of the network to which it is connected Make sure that the power cable does not touch the heating plate The tool can only be opened by specialized technicians Keep the instr...

Страница 2: ...ectively Power on overload led The LED can take on the color Green or Red When the instrument is on and in normal state the LED is green while it is red when overload protection is activated When the torque reaches the limit value the overload protection function is activated At the same time the light protection light flashes and the system stops Spindle For fixing the accessory through hole Thro...

Страница 3: ...cous i e an overload occurs the motor power is automatically regulated electronically When the torque limit value is reached the overload protection function is started and at the same time the red LED starts flashing to indicate a small overload La protezione contro il sovraccarico interviene nei seguenti casi When the value of the set speed is not suitable for the current average viscosity When ...

Страница 4: ...the instrument back in its initial packaging to send it to the repair service Reference standards The instrument was built in compliance with the following safety regulations EN 61010 1 UL 3101 1 CAN CSA C22 2 1010 1 EN 61010 2 10 The instrument has been manufactured in compliance with the following EMC standards EN 61326 1 European guidelines EMC guidelines 89 336 EWG Machine guidelines 73 023 EW...

Страница 5: ...ol cannot be disposed of in public landfills In accordance with EU directive 2002 96 EC European users of electrical and electronic equipment have the option of returning the used equipment to the Distributor or Manufacturer when purchasing a new one Abusive disposal of electrical and electronic equipment is punished with a pecuniary administrative sanction ...