The software can be found in menu Tools->”Pt100 Calibration Tool” of
the logger client program. Below you find a screen shot of the tool.
In this form you set the following fields:
Current gain:
The current gain parameter setting used to measure the reference
temperatures.
Current offset:
The current offset parameter setting used to measure the reference
temperatures.
Reference Temperature 1:
The lower reference temperature value in degrees Celsius
Measured Temperature 1:
The measured temperature value in degrees Celsius
Reference Temperature 2:
The upper reference temperature value in degrees Celsius
Measured Temperature 2:
The measured temperature value in degrees Celsius
When the parameters are set, you find in the two lower fields the para-
meters to set in the PRO-PT100 device:
Calculated gain as parameter 1.
The calculated offset as parameter 2.
After these parameters are set, you can verify if the sensor is indicating
the correct temperature for each reference. If needed you can repeat
the same procedure again.