RE. ENR.110 - Rév B – le 15/03/2014
page : 14
Procedure for pressure generator models CH2-1250B and CH2-2000B,
CH2-4000B and CH2-7000B and CO-800B and CO-1000B
> Hydraulic diagram
>
Pressure mounted
> Observe the maximum operating pressures
of the pressure generator
Maximum operating pressure : C0-800B
850 bar and CO-1000B
1050 bar
CH2-1250B
1270 bar and CH2-2000B
2020 bar and CH2-4000B
4020 bar
and CH2-7000B
7000 bar
a)
Connect the instrument to be tested (G1/2<1200bar / M16x150>1200 bar)
b)
Check the valves
A
and
B
are open
c)
Check the capstan
1
pressure generator is in the end position.
d)
Close the valve
A
, turn the capstan
1
in the direction of clockwise : you can build up pressure
up to 120 bar (Maximum operating pressure : 150 bar). Make readings of measurements points in
this measuring
> To generate higher pressures, close the valve
B
and open the valve
A
, turn the capstan
1
in
the direction of clockwise.
e)
Realize the measurement point.
>
Reduce Pressure
f)
After the measurement point readings, turn the capstan
1
in the opposite direction of
clockwise until the stop end position and open the valve
B
.
g)
The instruments can be disassembled.