p11
If validation reading of the µconsole file
(see page 16)
during
1s.
if transfer
correct
if transfer
incorrect
LOAD
4.9.3 Menu
: Teleloading of programming files
READ
CONS
COPY
µCo
µCo
µCo
µCo
LOAD
WAIT
OK
or
ERR
PROG
OPT
ERR
PROG
Saving of the converter programming into the µconsole
memory.
Teleloading of the programming file contained in the µconsole
to the converter.
Programming file incorrect: belongs to another type
of instrument.
The file does not correspond to the same options as
the instrument currently connected.
ERR
ERR
PROG
OPT
M
Reminder:
If no key is pressed during 20 s., or if is pressed the
instrument will revert to the measure display.
The file stored in the µconsole can be read by validating choice
READ
CONS
. In this case the organigramme for reading the programming of
the file is the same as for the programming of the instrument
(see p5).
5.1 Functions which require pressing only 1 key:
a / Display of the min. value
b/ Display of the max. value
INF
02.00
Measure
display
Alternating infor-
mation display
Min. value
5. FUNCTIONS WHICH CAN BE ACCESSED ON
THE FRONT FACE
SUP
08.00
Measure
display
Alternating infor-
mation display
Max. value
c / Deleting of the min. and max. values
CLrM
Measure
display
deleting of the recorded min.
and max., and revert to the
measure display
M
The instrument will revert to the measure display.
Reminder:
If no key is pressed during 20 s., the
instrument will revert to the measure display.
Note:
An exit from mode programming with saving of the confi-
guration will reset the min. and max. values to 0.