Arctiko ULUF 120 Скачать руководство пользователя страница 15

 

Description 

 

Arctiko default 

settings 

 

Customer 

settings 

Custom Settings        (Password) 

0000 

 

 

 

 

Set point 

 

 

Freezer 

 

 

 

 

-80,0

Power failure

 

 

 
 

 

 

 

High temp. alarm

 

 

 

Freezer 

 

Alarm delay

15 min

 

 

 

Low temp. alarm

 

 

 

Freezer 

 

 

 

 

Door open alarm 

 

 

Freezer 

 

-70

 

 

 

Probe/eprom failure

 

 

 

Freezer 

 

-90

 

 

 

 

 

Freezer 

 

 

 

 

 

 

Freezer 

 

Enable

 

 

 

 

15

Default settings ULUF 450/450-2M/550/550-2M/750/750-2M/850/850-2M:

 

Description 

 

Arctiko default 

settings 

 

Customer 

settings 

Advanced Settings    (Password) 

0000 

 

 

 

 

Hysteresis 

 

 

Freezer 

 

 

 

 

0,0

 

 

 

Freezer 

 

 

 

 

0

 

 

 

Freezer 

 

Manual defrost

 

 

 

 

 

 

Freezer 

 

 

 

 

Calibration 

 

 

2,0

Freezer  Max. 

 

Temp. range limits

 

 

 

-40,0

Automatic defrost

 

 

  

 

 

 

 

Alarm Settings:

Logging time interval

1

Enable

Freezer 

Enable

OK

Freezer  Min. 

-85,0

Содержание ULUF 120

Страница 1: ...OPERATING INSTRUCTIONS ULUF Models...

Страница 2: ...eezer Electrical connection Before commissioning Starting procedure Starting up the freezer Disconnecting the appliance Control Panel Default settings The cooling system CO2 back up Relocating or movi...

Страница 3: ...appliance contains a mixture of various refrigerants For more information please refer to the rating plate affixed to the freezer Power Failure The unit is equipped with battery back up for display an...

Страница 4: ...ree to the sides at least 15 cm free at the back Electrical connection Trouble free operation this unit must be properly grounded before it is used Failure to ground the equipment may cause personal i...

Страница 5: ...w water to come into contact with the cabinet Water entering electrical parts may cause short circuiting 6 In the event of functional or electrical irregularities please contact an authorised service...

Страница 6: ...4 5 6 7 8 USB Data download 1 Temperature indicator 2 Alarm indivator 3 Alarm icons 4 Compressor indication 5 Date indication 6 Hour indication 7 Ambient temperature indicator 8 Battery level 9 USB C...

Страница 7: ...7 Setting icons Custom Settings Advanced Settings Advanced Service Settings Status Change Reset password Change Reset password Status Advanced Service Settings Advanced Settings Custom Settings...

Страница 8: ...arm will occur on the unit and until it will be shown on the display The menu High temp alarm is the settings of the highest temperature the device must be in side before it comes with an alarm The me...

Страница 9: ...defrost the deftrost can be started immediately The menu Hysteresis is the setting of the different from the setpoint whwere the compressor will start and stop The menu Temp range limits is the settin...

Страница 10: ...splayed Under the menu Probe inside unit the tempe rature of the sensor mounted inside the unit is displayed There is mounted one sensor as stan dard in the unit but it is possible to mount 3 sensors...

Страница 11: ...ossible to change the password for Customer Settings Advanced Settings and Advanced Service Settings Under Reset password it is possible to reset the pass word for the Customer Settings Advanced Setti...

Страница 12: ...the display USB connected Download Data Place the USB pen drive in the connection on the display and press the button right for read data from the unit to the USB pen The Read data will be shown in t...

Страница 13: ...unit The Low temperature alarm indicate that the temperature inside the unit is lower than permitted from setting of the unit The Probe eprom failure alarm indicate that the probe is not working The...

Страница 14: ...er 30 Probe eprom failure Freezer 50 Freezer Freezer Enable 14 Default settings ULUF 60 120 400 500 700 800 Description Arctiko default settings Customer settings Advanced Settings Password 0000 Hyste...

Страница 15: ...e eprom failure Freezer 90 Freezer Freezer Enable 15 Default settings ULUF 450 450 2M 550 550 2M 750 750 2M 850 850 2M Description Arctiko default settings Customer settings AdvancedSettings Password...

Страница 16: ...m Freezer 70 Probe eprom failure Freezer 90 Freezer Freezer Enable 16 Default settings ULUF 15 65 125 Description Arctiko default settings Customer settings AdvancedSettings Password 0000 Hysteresis F...

Страница 17: ...m Freezer 70 Probe eprom failure Freezer 100 Freezer Freezer Enable 17 Default settings ULUF 490 590 890 Description Arctiko default settings Customer settings AdvancedSettings Password 0000 Hysteresi...

Страница 18: ...from the top of the unit remove kit Make sure that pressurized CO2 escape from the gaskets or pressure valve Relocating or moving the freezer Never leave the freezer to others without instructing thes...

Страница 19: ...cabinet and the evaporator system The compressor compartment and in particular the condenser must be kept free from dust and dirt This is best done with a vacuum cleaner and a brush Do not put bottle...

Страница 20: ...20 Wiring diagram ULUF 400 450 450 2M 500 550 550 2M 700 750 750 2M PT1000 PT1000 T3 T2 NTC NTC PT1000 100 240V AC 800 850 850 2M...

Страница 21: ...21 Wiring diagram for ULUF 15 60 65 120 125 PT1000 PT1000 NTC T3 T2 PT1000...

Страница 22: ...22 Wiring diagram for ULUF 490 590 890 T3 T2 PT1000 PT1000 PT1000 100 240V AC...

Страница 23: ...23 Notes...

Страница 24: ...50 80 038 10 GB WWW ARCTIKO COM LAMMEFJORDSVEJ 5 DK 6715 ESBJERG N DENMARK TEL 45 70 20 03 28 FAX 45 70 20 03 29 INFO ARCTIKO COM WWW ARCTIKO COM...

Отзывы: