5
Regulations require that this appliance must be earthed:
The manufacturer declines all liability for injury to persons or animals and for damage to property
resulting from failure to observe the above procedures and reminders.
If the socket and plug are not of the same type, please have the plug or socket replaced by a qualified
electrician.
Do under no circumstances use extension leads or adaptors.
1. The freezer should be connected by an authorised electrician or by the company, which sold the
freezer! Incorrect installation may damage the freezer.
2. Place the freezer on a flat floor, which can take the weight of the freezer.
3. The electrical installation should be earthed.
4. Do not use extension cables, but connect the freezer directly to a fixed installation.
5. The freezer should be placed in a dry location, and should not be subjected to water on the cabinet.
If water gets into the electrical parts it may cause short-circuiting.
6. In the event of functional or electrical irregularities, please contact an authorised servicing engineer.
Before Commissioning
Clean the freezer both inside and out using a damp, wrung out cloth. Dry with a dry cloth.
The smell from the plastic parts in the freezer will disappear once the freezer has been cooled down.
Start procedure
Put the plug in the socket.
If the battery voltage is low it will take about 2h before the compressor starts.
The thermostat was pre-set by the producer. The freezer should run for at least 6 hours before the
Starting up the freezer
Start the freezer. Set the digital thermometer at the temperature required. It will now take about 6-10
hours before the set temperature is reached. The freezer should be cooled down without anything in it.
Disconnecting the appliance
Electrical disconnecting must be possible either by unplugging the appliance or by means of a double
pole switch located up-line from the socket.
thermostat is changed (Stabilisation of the cooling system).
WARNING
Содержание ULTF 220
Страница 1: ...OPERATING INSTRUCTIONS ULTF 80 220 320 420...
Страница 16: ...16 Wiring diagram ULTF 80 Control box T3 T2 PT1000 PT1000 NTC NTC PT1000 PT1000 240V AC 100...
Страница 17: ...17 Wiring diagram ULTF 220 Control box T3 T2 PT1000 PT1000 PT1000 PT1000 NTC NTC 240V AC 100...
Страница 18: ...18 Wiring diagram ULTF 320 420 Control box Fan V PT1000 PT1000 PT1000 PT1000 NTC NTC T3 T2 240V AC 100...
Страница 19: ...19...