![Arbiter Systems 1073A Скачать руководство пользователя страница 21](http://html1.mh-extra.com/html/arbiter-systems/1073a/1073a_operation-manual_2962813021.webp)
3.2 Rack Mounting
9
then remove the cord from the rear-panel IEC connector. The in-circuit fuse is the innermost one;
inspect it to determine whether it is open. As required, replace with fuse in the outer compartment.
For instruments supplied with Option 08 or 10, the fuse is located in the fuse holder on the rear
panel near the power inlet terminal strip. The fuse is a 1-Amp, 250-Volt fast-acting (Option 10) or
time-delay (Option 08) 5 x 20 mm fuse. No spare fuse is provided for Options 08 or 10.
Replacement Fuses
Power Option
Arbiter P/N
Standard P/N
Option 07
FU0001816
F1AL250V
Option 08
FU0001419
T2AL250V
Option 10
FU0001816
F1AL250V
3.1.4
Rear-Panel Layout
The rear panel of the Model 1073A is arranged in the following manner, left to right (see Figure 3.1):
•
Five BNC type connectors, corresponding to channel C input and outputs.
•
Five BNC type connectors, corresponding to channel B input and outputs.
•
Five BNC type connectors, corresponding to channel A input and outputs.
•
One opening for an optional fiber-optic input connector (Option 01). If not used, it is covered
by a plastic hole plug.
•
Option 07 provides an IEC-320 power inlet connector with built-in fuse holder. Options 08
or 10 provide a 3-pole-terminal strip and separate fuse holder.
Figure 3.1: Model 1073A Rear Panel
3.2
Rack Mounting
To facilitate mounting the instrument in a standard 483 mm (19 inch) equipment rack, the 1073A
includes a pair of rack-mounting ears. To install the rack mounting ears:
1. Using a T-25 driver, remove the two screws on one side of the unit. Leave the cover in place.
2. Position one of the rack mounting ears on the side of the unit, so that the rack mounting
flange is at the front of the instrument and extends away from the front panel.
3. Replace the cover screws by routing them through the lower set of holes in the rack-mounting
ear, and back into the threaded holes in the instrument.
4. Repeat the above steps for the opposite rack-mounting ear.
Содержание 1073A
Страница 4: ...iv ...
Страница 6: ...vi ...
Страница 10: ...x LIST OF FIGURES ...
Страница 12: ...xii LIST OF TABLES ...
Страница 18: ...6 Technical Specifications Operational Parameters ...
Страница 26: ...14 Physical Configuration Figure 3 2 Model 1073A Main Board Jumper Locations ...
Страница 28: ...16 Operation ...