5. Installation and Operation

5-1. Power Connection

  Unpack the box, and check for the included components;

   A main body with an angle rotor coupled, a spanner, power cable, 
   and user manual.

  Do not bend or pull the power adaptor too hard.

  For the first time the power is connected, the lid closure will not 

  initiate spinning even if the lid is completely closed. It is for user’s
  safety. It will be properly functioned when the lid is simply open
  and closed.

1. Check if the rotor and the lock nut are on place and tightly fixed to the motor shaft. 

2. Connect the DC adaptor to the connector on the back of the main body. 

3. Plug in the power code into proper wall outlet.  

4. Small beeping sound will be heard when the power is correctly connected.

WARNING
Do not close the lid unless the rotor is property balanced with identical tubes.
Do not touch the rotor or tubes unless it makes a complete stop.

Do not use any other than the power adapter provided by GYROZEN.
(Adaptor INPUT : AC 100~240V, OUTPUT : DC12V)

Содержание mini 6

Страница 1: ...ARA mini 6...

Страница 2: ...RESEARCH USE ONLY PRODUCT INFORMATION Model ARA mini 6 Manufacturer GYROZEN Co Ltd 1F 5F 16 Arayuk ro Gimpo si Gyeonggi do KOREA ARA mini 6 UM mini 6 Rev 3...

Страница 3: ...e notice 1 Safety Precautions 4 1 1 Precautions for Installation 4 1 2 Precautions for Use 5 2 Product Description 5 3 Configuration 6 4 Product Specifications 7 5 Installation and Operation 8 5 1 Pow...

Страница 4: ...cautions for Installation 1 Avoid hot humid places and install in well ventilated areas 2 Keep a distance of 30 cm from the wall and install it horizontally on a vibration free rigid table 3 For safet...

Страница 5: ...hem symmetrically 4 Do not move the equipment while it is spinning 5 Do not arbitrarily decelerate or stop the rotating rotor 6 Do not spill liquid or put foreign substances into the body or power ada...

Страница 6: ...de Selection Switch Operation change switch Lid As soon as the lid is closed the rotor starts to spin As soon as the lid is opened the rotor stops to spin Short mode S Spin for 12 sec while the lid is...

Страница 7: ...e 2 234 xg 2 PCR Strips Rotor 1 412 xg Radius Angle Rotor 1 5 2 0 mL x 6 hole 47 3 mm 2 PCR Strips Rotor 29 9 mm Tube capacity 6 x 1 5 2 0 mL 2 x 8 tube PCR Strips Time Selectable Short 12 sec or Cont...

Страница 8: ...It is for user s safety It will be properly functioned when the lid is simply open and closed 1 Check if the rotor and the lock nut are on place and tightly fixed to the motor shaft 2 Connect the DC a...

Страница 9: ...5 2 Operation 1 Rotor coupling and decoupling This model is supplied with a coupled rotor In case of any needs of decoupling use the supplied spanner Counterclockwise turns make decoupling and clockw...

Страница 10: ...2 Max RPM RCF 6 500 rpm 1 379 xg 6 500 rpm 1 710 xg 6 Trouble Shootings If the equipment does not operate 1 Incorrect connection between power adaptor and equipment Please check if the main body and t...