(*) If there is data in the logger memory, parameters marked with (*) cannot be opened.
If they should be changed, the data must first be deleted!
Menu
Parameter
OFF
Standard hold function at the push
of a button (only with Logger = oFF)
see
A
VTO
HLD
1 - 365
Time interval for calibration reminder
(in days)
OFF
No calibration memory
P.OFF
1 - 120
Auto power off
(shut-off delay) in
minutes. If no button is pressed and
data exchange no longer takes place
over the interface, the device shuts off
automatically after the lapse of this time
(factory setting: 20 min)
OFF
Automatic shut-off deactivated (contin-
uous operation)
L,TE
OFF
No background lighting, lowest
power consumption
5 - 120
Automatic shut-off of lighting after
5 - 120 s (factory setting: 5 s)
ON
Background lighting always on
0V T
OFF
No output function, lowest power
consumption
SER
Device output is serial interface
(factory setting)
DA[
Device output is analog output
ADR
0 1, 1 1,2 1, …
9 1
Base address of the device for serial
interface communication (factory
setting 01)
DA.,N
[ONC
Analog output based on
concentration in ppm or mg/l
SAT or P 02
Analog output based on setting of
(H 2
DA[.0
0 . 0 - 1 0 0 . 0
%O2
Input of O2 concentration at which
the 0V analog output should be
output (factory setting 0.0 %O
2
)
DA[. 1
0 . 0 - 1 0 0 . 0
%O2
Input of O
2
concentration at which
the 1V analog output should be
output (factory setting 100.0 %O
2
)