AL400T-WL_2a 07/2011
1.4.2 Cleaning of vials & sampling containers
Vials, caps and sampling containers should be cleaned thoroughly after each test to avoid
influences. Minor residuals can cause errors.
Residuals:
Cleaning of the vials will vary according to the different types of water sample used.
Replace scratched vials immediately.
•
Rinse vials thoroughly with deionised water after each measurement.
•
Clean all glassware thoroughly with laboratory detergent and rinse with deionised water.
•
Clean heavy contamination by filling the vials with 1:1 HCL followed by multiple rinses
•
with distilled or deionised water.
Allow vials to air dry.
•
Touch vials only at the top to minimise dirt and fingerprints.
•
Wipe the vials with a lint-free cloth to remove waterdrops and fingerprints.
•