Dip the probe
in the water
Waiting
Rx 475 mV
Probe Quality
4) after 5 seconds, it displays the value of the measured chlorine (uncalibrated) from the probe and
expressed in ppm.
5) If you press the Enter key returns to the value of the previously measured chlorine. Use the Up and
Down keys to change this value to the value measured with the photometer (DPD method).
Pressing for about a second the Enter key to return to the installer menu item Calibration Chlorine.
4.5.4
Redox
probe
calibration
menu
The Redox Probe Calibration Menu calibrates the Redox probe in case the model supports it. Probe
calibration regulates the measurement tool and improves its
efficiency.
It must be performed before using the
probe, following the step-by-step procedure below:
1) Rinse the probe and dip it in the buffer solution at 475mV, move slowly for 10 seconds and press
Enter.
2) Wait without touching the probe or its cable for at least 60 seconds, until the value read is
stabilized.
3) At the end of the countdown or after pressing Enter, the Redox value of the buffer solution (475
mV) and the value read by the probe will be shown.
4) The procedure ends visualizing the quality of the probe with a percentage from 0 to 100. If quality
is below 25%, we recommend to replace the probe.
5) Place the probe in the probe holder or in the filtration system.
Press
Enter
to scroll the menu sections.
Press and hold Enter for about 1 second to go back to the Redox Calibration section of the Installer Menu.
4.5.5
pH probe
calibration
menu
The pH Probe Calibration Menu calibrates the pH probe in case the model supports it. Calibration regulates
the measurement tool and improves its
efficiency.
Calibration may have one or two reference points. pH
probe calibration must be done as follows:
1) Rinse the probe in a bottle of drinking water
2) Dip the pH probe in the buffer solution (pH 7), slowly move for 3 seconds and press
Enter.
3) Wait without touching the probe or its cable for at least 60 seconds, until the value read is
stabilized, then press Enter.
4) At the end of the countdown or after pressing Enter, the pH value of the solution (7.00) and the
value read by the probe (in mV) will be shown.
5) For a calibration with one reference point, press F and move to point 8. Otherwise, rinse the probe
in the drinking water bottle and dip it in the pH 4.01 solution. Slowly stir for 3 seconds, then press
Enter.
Dip the probe
in the water
Waiting
Chlorine mA
read chlorine
Set chlorine