- Install connection hoses as follows:

○ Push the red hose or existing ø8mm hose/pipe completely

into the quick release adapter (filter inlet side)

○ Push the blue hose completely into the quick release adapter

(filter outlet side)

○ Pull the pipes to check the connection

4.2 Filter installation  

- Disinfect fittings with an alcoholic surface disinfectant
- Open the packaging and remove the filter
- Connect the filter inlet to the red hose or existing ø8mm hose/pipe
- Similarly, connect the blue hose into the filter outlet side

- Pull the pipes to check the connection

-

Open water supply, check tightness of the connections

and water flow

4.3 Filter replacement documentation  

- Note the date of next filter replacement on the sticker and

place it on the filter in an easily visible position –

but

 not

 on the barcode.

5 Removal of used filter / filter change

Please disinfect your hands and wear disposable gloves during filter installation

Turn off general water supply before installation

The water in the pipes must not be under pressure

- Disconnect the blue hose from the filter

(Caution: Residual water can escape).

The connection is released by pressing
the retaining element against
the connector.

- Place zip lock bag over the filter
- Disconnect the red filter inlet hose (as above)

(Caution: Residual water can escape)

- Hoses remain attached to the adapters
- Place the filter into the provided transport box
- Equip transport box with protective foil for safe transport

- Close transport box with cable ties

- Install a new filter (see point 4.2)

4 Preparation of commissioning and use of product

  

Please disinfect your hands and wear disposable gloves during filter installation 

Turn off general water supply before installation

4.1 Adapter and hose installation

For initial installation or maintenance – if appropriate connection options (hose / pipe ø 8 mm) are not 
available

- Open the water line to the point where the filter is to be installed. Note: Residual water can escape.

- Install the suitable quick-release coupling on counterparts in the pipe system. Note: Pay attention the

correct seating of the seal.

- Cut back the connecting hoses to the appropriate length, keeping blue hose as short as possible

1.4 Reprocessing

The Germlyser

®

 IL 1500 is a high-quality, reusable        -labelled medical device that is made available to the 

customer at a fixed rental rate. Reprocessing is carried out exclusively by Aqua 

free

.

2 Certificates / Declaration of conformity

The company Aqua

 free

 GmbH is certified according to DIN EN ISO 13485. 

The Germlyser

®

 filters are       -labelled medical devices.

Extract from the declaration of conformity  
Aqua free

 GmbH, 

Winsberging

 3

1

, DE-22525 Hamburg:

The manufacturer declares for the product Germlyser

®

 IL 1500 to Annex IX of the directive 93/42 EWG 

on medical devices and medical products of Class I the conformity with the basic requirements of Annex 

I of the council directive 97/42 EEC of the European communities.

3 Labelling

Manufacturer label

D/GB

Product information material

SN

Individual product label

Medical device label

Non-sterile

Do not use if package 

is damaged

NON 

STERILE

1.3 Scope of delivery

• Germlyser

®

  IL 1500

• Transport box

• Cable tie for product return and box fastener
• Zip lock bag for used filter
• Product information, sticker for filter replacement