MEM
SET
START
STOP
Exceptional Situation
Exceptional Situation
Care and maintenance
Care and maintenance
Specification
Specification
About Blood Pressure
Care and maintenance
Warranty Information
EMC Declaration
EMC Declaration
Warranty Information
EMC Declaration
EMC Declaration
EMC Declaration
EMC Declaration
Table of Contents
Optimal blood pressure
Normal blood pressure
High normal value
Mild hypertension
Moderate hypertension
Severe hype rtension
Blood Pressure Type
Error indicators
Maintenace
Trouble removal
Symbol
Cause
Correction
Problem
No power
Err and stop
working
Cuff leak
Check
Check the battery power
Check the polarity position
Replace new one
Installation for proper placement
of the batteries polarities
Whether move the wrist when inflate Keep the body peaceful
Check if chatting when measured
Keep quite when measure
Whether the cuff wrap too loose
Wrap the cuff tightly
Whether the cuff broken
Please contact the distributor if you can't solve the problem, do not disassemble the
unit by yourself!
Change a new cuff
Cause and solutions
The following symbol will appear on the display when measuring abnormal.
Weak signal or
pressure change
suddenly
Wrap the cuff properly.
Remeasure with correct way.
When near cell phone or other high radiant device the
measurement will be failed.
Keep quite and no chatting when measure.
Wrap the cuff properly.
Make sure that the air plug is properly inserted in the unit.
Remeasure.
Repeat the measurement after relax for 30 mins,if get
unusual readings for 3 times,please contact your doctor.
Replace all the worn batteries with new ones.
External strong
disturbance
It appears error
during the
process of
inflating
Abnormal blood
pressure
low battery
Keep the unit in the storage case when no
use.
Clean the unit with soft dry cloth.
Do not use any abrasive or volatile cleaners.
Never immerse the unit or any component in
water.
Make sure the monitor is off prior to cleaning,a
mixture of distilled water and 10 percent bleach
could be used.
Using a spray bottle, moisten a soft cloth towel
with the bleach or detergent mix until it is fully
saturated. Squeeze any excess moisture from the
cloth to avoid any dripping or potential oversatu-
ration of the cuff.
Wipe all surfaces of the blood pressure monitor
cuff thoroughly, making sure to clean the inside
and outside of the cuff. Be cautious not to get any
moisture in the main unit.
Using a dry cloth, gently wipe away any excess
moisture that may remain on the blood pressure
cuff. Lay the cuff flat in an unrolled position and
allow the cuff to air dry.
We won' t be responsible for any quality problem if you don't care and maintain the
product as instructed.
Remove the batteries if the unit will not
be used in 3 months or longer.
Do not wet the cuff or attempt to clean
the cuff with water.
Do not clean the body and cuff with
naphtha, thinner or gasoline etc.
Store the unit in a clean and dry
location.
Do not subject the unit to extreme hot or
cold temperature, humidity and direct
sunlight.
2pcs AAA alkaline battery DC.3V
2x199 sets memory of measurement values
Memory/Heartbeat/Low battery
3 digits display
3 digits display of mmHg
±5% of reading
±3mmHg
40~199 pulses/min
0~299 mmHg
Symbol
Pulse
Pressure
Pulse
Pressure
Pulse
Pressure
wrist
Oscillometric method
LCD digital display
Automatic wrist blood pressure monitor
Power source
Memory function
LCD
indication
Accuracy
Measurement
range
Measuring loca ization
Measuring principle
Display
Description
L69.5mmXW66.5mmXH60.5mm
Approx.96g (batteries not included)
In 3 minutes
Could be used for 300 times for normal condition
Cuff, instruction manual
Temperature
Humidity
Air pressure
5~40°C
15% ~ 93%RH
80kPa ~ 106kPa
Temperature: -20°C ~ 55°C, Humidity: 10% ~ 93%
avoid crash, sun burn or rainduring transportation.
Main unit size
Main unit weight
Automatic power off
Battery life
Accessories
Operating
environment
Storage environment
Cuff size
Wrist circumference approx. Size 13.5~21 .5cm
The intended use: the unit is intended to be used by adults at home or medical
center to measure blood pressure and pulse rate from the wrist.
The unit satisfies the requirements of EN ISO 81060-1 Part 1 Noninvasive
sphygmomanometers, EN 1060-3:1997+ A2:2009 Non-invasive sphygmomanom-
eters. IEC80601-2-30 Part 2 Non-invasive sphygmomanometers.
Blood pressure measurements determined with this device are equivalent to those
obtained by a trained observer using
the cuff/ stethoscope auscultatory method, within the limits prescribed by the
American National Standard
,
manual,electronic, or automated sphygmomanome-
ters.
The risk of patient and user can be lowered to acceptable level.
The unit is guaranteed to be free of defects in workmanship and materials under
normal use for a period of One Year from the date listed on the purchase record.
For repair under this warranty. Our authorized service agent must be advised of the
fault with the period of the warranty. This warranty covers parts and labor only under
normal operations. Any defect resulting from natural causes, eg. flood,hurricane
etc, is not within this guarantee. This guaranty does not cover damage incurred By
use of the unit not in accordance with the instructions,accidental damage,or being
tampered with or serviced by unauthorized service agents.
Monitor subjected to misuse,abuse, and neglect of these manual content,non-in-
structional purposes;unauthorized repair or modifications will be excluded from this
warranty.
The device requires no calibration.
The device is not repairable and contains no user serviceable parts.
Statement
Warranty Information
IEC 60601-1-2: 2014 ME EQUIPMENT and ME SYSTEMS identification,
marking and documents for Class B product
Instructions for use
The ME EQUIPMENT or ME SYSTEM is suitable for home healthcare
environments and so on.
Warning:
Don’t near active HF surgical equipment and the RF shielded room
of an ME system for magnetic resonance imaging, where the intensity of EM
disturbances is high.
Warning:
Use of this equipment adjacent to or stacked with other equipment
should be avoided because it could result in improper operation. If such use is
necessary, this equipment and the other equipment should be observed to
verify that they are operating normally.
Warning:
Use of accessories, transducers and cables other than those
specified or provided by the manufacturer of this equipment could result in
increased electromagnetic emissions or decreased electromagnetic immunity
of this equipment and result in improper operation.”
Warning:
Portable RF communications equipment (including peripherals such
as antenna cables and external antennas) should be used no closer than 30
cm (12 inches) to any part of the blood pressure monitor, including cables
specified by the manufacturer. Otherwise, degradation of the performance of
this equipment could result.
If any:
A list of all cables and maximum lengths of cables (if applicable),
transducers and other ACCESSORIES that are replaceable by the
RESPONSIBLE ORGANIZATION and that are likely to affect compliance of t
he ME EQUIPMENT or ME SYSTEM with the requirements of Clause 7
(EMISSIONS) and Clause 8 (IMMUNITY). ACCESSORIES may be specified
either generically (e.g. shielded cable, load impedance) or specifically (e.g. by
MANUFACTURER and EQUIPMENT OR TYPE REFERENCE).
If any:
The performance of the ME EQUIPMENT or ME SYSTEM that
was determined to be ESSENTIAL PERFORMANCE and a description of
what the OPERATOR can expect if the ESSENTIAL PERFORMANCE is
lost or degraded due to EM DISTURBANCES (the defined term “ESSENTIAL
PERFORMANCE” need not be used).
Wrist Electronic Blood Pressure Monitor
※
We won't be responsible for any quality problem if you don't care and
maintain the product as instructed.
MEM
SET
START
STOP
MEM
SET
START
STOP
MEM
SET
START
STOP
26/41
25/41
27/41
28/41
30/41
29/41
31/41
32/41
34/41
33/41
35/41
36/41
38/41
37/41
39/41
40/41
Technical description
1.All necessary instructions for maintaining BASIC SAFETY and ESSENTIAL
PERFORMANCE with regard to electromagnetic disturbances for the excepted
service life.
2.Guidance and manufacturer’s declaration -electromagnetic emissions and
Immunity.
4