background image

 

 

 

 1/7 

 

 

Heater Jacket 

 
 

Model HJSA & HJSB 

 
 
 
 

Manual 

 
 

 
 

 
 
 
 
 
 
 
 
 
 
 
 
 

 

 
 
 

 

 
 

 

 

 

AQ M-Tech AB

 

Содержание HJSA

Страница 1: ...1 7 Heater Jacket Model HJSA HJSB Manual AQ M Tech AB ...

Страница 2: ...2 7 Heater Jacket Model HJSA HJSB Manual version 3 3 February 2017 AQ M Tech AB ...

Страница 3: ...arranty 4 Warning 4 Manufacturer information 4 2 Introduction 5 3 Installing Heater Jacket 5 4 Cleaning 5 5 Connections 5 6 Technical specification 5 7 Material selection 5 8 Dismantling the Heater Jacket 5 9 Dimensions and weight 6 10 Accessories Spare parts 6 11 Fault tracing 7 12 Dealer 7 ...

Страница 4: ...pair or replacement of other equipment in addition to the Heater Jacket can not be provided under the terms of this warranty The purchaser pays the carriage charges to AQ M Tech AB AQ M Tech AB can in no way be held liable by the purchaser for damage caused to any other equipment Warning The Heating Jacket may only be connected to one Heater Jacket Control Unit intended for this purpose Touching t...

Страница 5: ...eating cut out HJ1 155 C 155 C Maximum application temperature 140 C 140 C Maximum operating time at 140 C 400h 400h Temperature range SV 2 30 C 140 C 30 C 140 C Normal working temperature 30 C 100 C 30 C 100 C Temperature sensor type Pt 100 Pt 100 Protection class IP 65 IP 65 Ambient humidity 10 90 10 90 Steel material Stainless acid proof 316L Stainless acid proof 316L Plastic material Polyamide...

Страница 6: ... Millipore 10 138 302 104 1 6 400 115 HJSA 104 155 Millipore 5 138 179 104 1 4 200 230 HJSB 104 155 Millipore 5 138 179 104 1 4 200 115 HJSA 101 302 Sartorius Std 10 136 302 102 1 6 400 230 HJSA 101 162 Sartorius Std 5 136 162 102 1 4 200 230 HJSA 076 136 Sartorius mini 5 110 136 76 1 0 150 230 Series 3000 10 Accessories Spare parts Heater Jacket Control Unit Item Model Size WxLxH Item no Heater J...

Страница 7: ...lem Probable cause Corrective action 7 HJ does not become warm Power switch not pressed in on the HJCU No voltage Blown fuse Faulty cable connection HJ Short circuit in the HJ Faulty cable connection Temp HJCU or HJ faulty Press the power switch on the HJCU Check the incoming voltage Change the main fuse in the HJCU F1 Check the connection Contact the dealer Check the connection Contact the dealer...

Отзывы: