High-Power (P-HI) Mode Calibration
1.
At the FREQUENCY Display, press the
﹀
button to select
Π−ΗΙ
(P HI).
2.
At the OUTPUT Terminals, carefully connect a resistive load and connect a
certified, calibrated RMS power meter, of at least 0.2% accuracy.
3.
The size the load should be calculated based on the high-current parameter for
your particular APS unit, as shown in the following Calibration Table. Use 120
VAC for the voltage.
4.
An example calculation is shown on page 6-2 in the Section
Calculation Setup
,
subsection
How to Calculate the Size of the Load.
Calibration Settings
Parameter
1002
1003
1005 1010 1020 1030 1040
1060
V LO (VAC) 150.0
150.0 150.0 150.0 150.0 150.0 150.0
150.0
V HI (VAC)
300.0
300.0 300.0 300.0 300.0 300.0 300.0
300.0
A LO (A)
3.000
3.000 30.00 30.00 30.00 30.00 30.00
300.0
A HI (A)
16.00
25.00 42.00 80.00 160.0 250.0 330.0
500.0
P LO (kW) 0.3000 0.3000 3.000 3.000 3.000 3.000 3.000
30.0
P HI (kW)
2.000
3.000 5.000 9.000 18.00 27.00 36.00
54.0
5.
At the VOLTAGE Display, press the OUTPUT/RESET button. The APS unit
will automatically perform a high-power output reading and will output
approximately 120 VAC.
6.
The external RMS power meter will display the actual power output. The
CURRENT Display shows the calibration value (see Calibration Table, above).
7.
Adjust the load or the output voltage until the RMS power meter displays the
value shown in the Calibration Table (above). Please note, the value for your unit
will depend on the model.