CHAPTER 4
OPERATING INSTRUCTIONS
Entire Contents Copyright
2004 by Adaptive Power Systems, Inc. (APSI) • All Rights Reserved • No reproduction without written authorization from APSI.
APS-1000 Series
4 - 7
OM-001-01000-000-04.0
Power / PF (Display) Setting
When the system is in RESET (Standby) or the TEST (Output ON) mode, you can
switch between Power and Power Factor (PF) mode. However, unless the system is in
TEST (Output ON) you will not have a displayed value.
Power / PF Display P and PF Buttons
Use the button beneath the P / PF indicators to switch between modes.
Power / PF Display M1, M2, and M3 Buttons
These buttons are used to command the system to store (program) a set of values:
voltage, frequency, and current limit. The buttons are also used to recall the stored set
of values for output, during a test.
During RESET (Standby), press and holding "down" a button to command the system
to store a set of values. When you first depress the button, The related indicator begins
to flash. When the flashing stops, release the button and the values will be stored in
that memory set.
During TEST (Output ON), press and release a button to command the system to output
a set of stored values.
For more detailed information about the M-buttons, see the Section
Program Memory
Procedures
.
P O W E R / PF
M1
M2
M3
P
PF