Whether you are performing all calibrations or just a subset, perform calibrations for
the QuantStudio
™
6 and 7 Flex Systems in the following sequence.
Note:
A normalization calibration is not required for the 96-well plate blocks (0.1-mL
or 0.2-mL) or the 384-well plate block.
Start
▼
▼
Background calibration (page 45)
▼
Uniformity calibration (page 49)
▼
▼
Normalization calibration (TaqMan
▼
Verify the instrument performance (page 62)
(RNase P instrument verification)
▼
Finish
You can use the Instrument Console of the QuantStudio
™
Real-Time PCR Software to
view the calibration status of an instrument.
1.
In the QuantStudio
™
Real-Time PCR Software Home tab, click Instrument
Console.
2.
From the Instrument Console open
the Instrument Manager.
a.
Select your instrument icon
from the My Instruments
group.
b.
Click Manage Instrument.
Note:
If the Manage Instrument
option is inactive, connect your
instrument to the My Instruments
group.
Calibration
workflow
View the
calibration status
of an instrument
Chapter 2
Calibrate the instrument
Calibration overview
2
QuantStudio
™
6 and 7 Flex Real-Time PCR Systems v1.6 Maintenance and Administration Guide