5-8 Creating and Editing Methods
Entering
Temperature
Control Parameters
When you enter temperature control parameters, you define values for parameters in
each of the three segments of a method: pre-PCR, PCR, and post-PCR.
Figure 5-1
Order of advancement of the highlight box
To enter temperature control parameters:
Step
Action
1
On the
Create
screen, select a field.
When you first display the Create screen, the
Hld
field is highlighted.
2
Use the numeric keys to enter values.
3
Press
Enter
to accept a value. The next field is then selected in the order shown in
Number
Description
1
Number of pre- PCR holds
2
Pre-PCR temperature parameter
3
Pre-PCR hold time parameter
4
Number of PCR segment temperatures
5
Number of cycles for the PCR segment
6
PCR segment temperature parameter
7
PCR segment time parameter
8
PCR segment temperature parameter
9
PCR segment time parameter
10
PCR segment temperature parameter
11
PCR segment time parameter
12
Number of post-PCR holds
13
Post-PCR temperature parameter
14
Post-PCR hold time parameter
15
Post-PCR temperature parameter
16
Post-PCR hold time parameter
1 Hld
3 Tmp
25 Cycles 2 Holds
94.0
94.0
5:00
0:30
55.0
0:30
72.0
0:30
72.0
7:00
4.0
F1
F3
F4
F5
F2
Start
Cancel
More
Store
Pre-PCR segment
PCR segment
Post-PCR segment
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
Содержание GeneAmp PCR System 9700
Страница 1: ...GeneAmp PCR System 9700 Base Module User s Manual ...
Страница 9: ...ix To Obtain Customer Training Information C 7 D Limited Warranty Statement Index ...
Страница 10: ......
Страница 28: ......
Страница 56: ......
Страница 90: ......
Страница 96: ......
Страница 110: ......
Страница 116: ......
Страница 124: ......
Страница 126: ......
Страница 131: ......