4.
Click
Discover
in each specific application window to identify the instrument that you want to
connect to.
Figure 16 Parallel ATC instruments control and monitoring through multiple instances of ATC Demo
software
Note:
We recommend that you use one computer to control multiple ATC instruments, and we do
not recommend that a single ATC is connected to multiple computers.
Firmware upgrades
1.
Copy the
atc-X.X.X.update
firmware file into the USB thumb drive and plug this into the ATC
control box USB port.
2.
Connect the ATC to the ATC demo software as described in“Operate ATC as stand alone using
computer”
3.
Click
Upgrade
.
Chapter 2
Operate the Automated Thermal Cycler (ATC) System as a stand-alone system
Firmware upgrades
2
Automated Thermal Cycler (ATC) System User Guide
23