Maintaining the Instrument
Recommended Maintenance Schedule . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 107
Archiving and Backing Up SDS Files . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 109
Decontaminating the Sample Block
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 110
Moving the 7300/7500/7500 Fast System . . . . . . . . . . . . . . . . . . . . . . . . . . . . 117
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 120
Updating the Operating System Software and Service Packs
Recommended Maintenance Schedule
Weekly
Maintenance
Tasks
• Check disk space
• Archive or back up SDS plate document files (see
)
• Cycle the computer and instrument power (power off, then power on the computer
and instrument)
• Wipe instrument surfaces with a lint-free cloth
IMPORTANT!
Never use organic solvents to clean the 7300/7500/7500 Fast system.
Monthly
Maintenance
Tasks
• Perform a background calibration (see
)
• Perform an optical calibration on 7500/7500 Fast systems (see
• Clean up and defragment the computer hard drive (see
Su M T W Th F S
Week (7 Days)
Su M T W Th F S
Month (30 Days)