Appendix A
Moving the 7300/7500/7500 Fast System
Applied Biosystems 7300/7500/7500 Fast Real-Time PCR System Installation and Maintenance Guide
115
Notes
A
Moving the 7300/7500/7500 Fast System
PHYSICAL INJURY
HAZARD.
Do not attempt to lift the instrument or
any other heavy objects unless you have received
related training. Incorrect lifting can cause painful and
sometimes permanent back injury. Use proper lifting
techniques when lifting or moving the instrument. At
least 2 people are required to lift the 7300/7500/7500
Fast instrument.
IMPORTANT!
Moving your Applied Biosystems
7300/7500/7500 Fast Real-Time PCR System can
create subtle changes in the alignment of the
instrument optics.
1.
Load the packing plate or empty 96-well plate in
the instrument.
2.
Manually raise the block: from the ROI Inspector
window:
a.
Click
(or select
File New
).
b.
In the New Document wizard,
click
.
c.
In the SDS software, select
Instrument Calibrate
.
d.
In the warning dialog box, click
to
lower the sample block.
The ROI Inspector dialog box opens.
e.
In the ROI Inspector dialog box,
click
.
3.
Move your 7300/7500/7500 Fast system
according to the guidelines on
.
4.
Connect the components of the system (see
“Setting Up the 7300/7500/7500 Fast
Instrument” on page 20
).